5.4. Shifting to the Parameter Setting Mode
5.5. Condition Data Setting Procedure
This section describes an example of setting the number of moving
averaging times as a typical example. Take the same steps for other
parameters.
5.6. Scaling Data Setting Procedure
This section describes the concept of scaling data and an example
of setting a full-scale indicated value as a typical example. Take the
same steps for other parameters.
* The digital limiter is the function that limits an input value to the
relevant set value even if a signal higher than or lower than the
indicated value set to DLHI or DLLO respectively is input. The
setting condition is DLHI > DLLO, and if this condition is not met,
Err5 is caused, returning the display to DLHI.
Condition data
Linearize data
Measurement operation
Pressing for a prolonged time
Comparator data
Scaling data
Shift data
Pressing the Enter key causes the instrument to store data
and then return to measurement operation
①
Press the Enter key for approx. 3 seconds during
measurement operation.
②
Press the Mode key to move to the number-of-averaging
times setting.
④
Press the Mode key to move to the number-of-moving
averaging times setting.
⑤
Press the Increment key several times to set a desired
value (this setting example sets the number of moving
averaging times to "4").
Measurement operation
Pressing for a prolonged time
or
③
Press the Increment key to move to moving averaging
setting.
⑥
Press the Enter key to return to measurement operation.
(If the Mode key is pressed, you will be returned to the
setting menu; press the Increment key to move to the
next parameter.)
Input voltage : 0 to 700 V
0 to 700.0
:
Display
F S C : 7000
F I N : 9999
O F S : 0
O I N : 0
D L H I : 9999
D L L O : -9999
D E P : Lights up at the 10
1
place.
RANG : 15
* For 15 ranges, the full-scale
input value (FIN) is "9999" when
700 V is input.
Display
Input
7000
700V
0
Input signal : 0 to 1 mA
0 to 5000
:
Display
F S C : 5000
F I N : 9999
O F S : 0
O I N : 0
D L H I : 9999
D L L O : -9999
D E P :
RANG : 22
None
Display
Input
5000
1mA
0
Instruction Manual for MODEL A7X14,5,6-X(AC Voltage and Current Measurement) 6/12
Input voltage : 2.5 to 10 V
2.000 to 5.000
(display range:2.000 to 5.000)
:
Display
F S C : 5000
F I N : 9999
O F S : 2000
O I N : 2500
D L H I : 5000
D L L O : 2000
D E P : Lights up at the 10
3
place.
RANG : 13
Display
5.000
10V
0
DLLO
2.000
2.5V
DLHI
Input
Lights up at the 10
3
place.
Input signal : 0 to 5 A
0 to 5.000
:
Display
F S C : 5000
F I N : 9999
O F S : 0
O I N : 0
D L H I : 9999
D L L O : -9999
D E P :
RANG : 26
* For 26 ranges, the full-
scale input value (FIN) is
"9999" when 5 A is input.
Display
Input
5000
5A
0