28
Start-up
Functions control units
Proxeo Ultra
PB-510
PB-520
PB-530
Rinsing function for automatic internal cleaning of the coolant channels
✔
✔
✔
Cleaning function for automatic internal cleaning of the coolant channels
–
–
✔
> Before every patient: Perform rinsing funcion for automatic internal cleaning of the coolant channels.
> W&H recommends performing the cleaning function once a month using the active substance chlorhexidine 0.2%.
Fill coolant tank with at least 200 ml cleaning liquid. Then perform a rinsing function with tap water.
Approved coolants and rinsing liquids
> Physiological saline solution (NaCl, 0.9%)
> Hydrogen peroxide (H
2
O
2
, 1–3%)
> Liquids with the active substance chlorhexidine (CHX, 0.2%)
> Tap water
Approved cleaning agents
> Liquids with the active substance chlorhexidine (CHX, 0.2%)
W&H recommends performing a rinsing function with tap water after using one of the approved liquids.