Hygiene and maintenance
> Do not place the handpiece with cable, the universal support, the stand and the instrument changer in liquid disinfectant
or in an ultrasonic bath.
Handpiece with cable / Universal support / Stand / Instruments / Instrument changer
> Clean the handpiece with cable, the universal support, the stand, the instruments and the instrument changer under
running tap water (< 35 °C / < 95°F).
> Rinse and brush off all internal and external surfaces.
> Remove any liquid residues using compressed air.
Control unit
> Do not immerse the control unit in water or clean it under running water.
Instruments
> Clean and disinfect diamond coated instruments in an ultrasonic bath.
Evidence of the basic suitability of the instruments for effective manual cleaning and disinfection was provided by an
independent test laboratory using the “Bandelin Type RK 100 CC“ ultrasonic bath and the cleaning agent and disinfectant
“Stammopur DR8“ (DR H Stamm, Berlin) and “CaviCide™“ (Metrex).
Manual cleaning