Operation
42
Waltron User Manual
3051 Hardness Analyzer
6.5
ONLINE MEASUREMENT
In online mode, the analyzer will run perform consecutive analysis cycles until it is manually
stopped or the out of sample alarm is tripped.
NOTE:
When starting the analyzer for the first time or restarting it
from any period of shutdown, do not proceed to put the
analyzer into online operation until a calibration has been
performed.
1.
Connect the sample tubing to the fast-loop reservoir and turn on sample flow to the fast
loop reservoir.
2.
Check that the sample fills the fast loop reservoir so that the float switch rises and sample
is being discharged from the overflow drain.
3.
Put the analyzer into online measurement mode by pressing the RUN Menu and selecting
START ON LINE.
The absorbance detected by the analyzer is used along with the blank and
calibration values to determine a concentration in ppb.
4.
When the analyzer has finished the analysis cycle, the concentration reading will be
displayed on the main screen. The new result and previous results are stored in the data
logger (up to 500 measurement results).
5.
The analyzer will wait the user defined amount of time (see Section 5.7.1) and then
proceed to run additional analysis cycles.
6.
To bring the analyzer out of online operation, press the RUN Menu. The top box where
START ONLINE was previously, will now be blank and dark gray. Pressing this box
will tell the analyzer to complete the current analysis cycle and go into stand-by mode at
its completion.
6.6
GRAB SAMPLE
The grab sample feature allows the user to measure the concentration of an external sample.
NOTE:
When starting the analyzer for the first time or restarting it
from any period of shutdown, do not proceed to run a grab
sample analysis until a calibration has been performed.
1.
If the analyzer is in online operation, it is necessary to put it into standby mode.
Press the RUN Menu and select the START ON LINE button (blank, dark gray box
on the top of the RUN Menu).
The analyzer will finish the analysis cycle that is currently in progress and go into
standby mode.
2.
Turn off sample flow to the fast loop reservoir.
3.
Disconnect the sample tubing from the fast-loop reservoir and connect it to the grab
sample source.
The grab sample source should have at least 200 ml of sample present.
4.
Perform an analysis cycle by pressing the RUN Menu and selecting CH 1.
Содержание AQUALERT Aqualyzer 3051
Страница 52: ...Appendix A Panel Mount Schematics 51 Waltron User Manual 3051 Hardness Analyzer ...
Страница 53: ...Appendix B Help Menu 52 Waltron User Manual 3051 Hardness Analyzer APPENDIX B HELP MENU ...
Страница 54: ...Appendix B Help Menu 53 Waltron User Manual 3051 Hardness Analyzer ...
Страница 55: ...Appendix B Help Menu 54 Waltron User Manual 3051 Hardness Analyzer ...
Страница 56: ...Appendix B Help Menu 55 Waltron User Manual 3051 Hardness Analyzer ...
Страница 57: ...Appendix B Help Menu 56 Waltron User Manual 3051 Hardness Analyzer ...
Страница 58: ...Appendix B Help Menu 57 Waltron User Manual 3051 Hardness Analyzer ...
Страница 59: ...Appendix B Help Menu 58 Waltron User Manual 3051 Hardness Analyzer ...
Страница 60: ...Appendix B Help Menu 59 Waltron User Manual 3051 Hardness Analyzer ...
Страница 61: ...Appendix B Help Menu 60 Waltron User Manual 3051 Hardness Analyzer ...
Страница 62: ...Appendix B Help Menu 61 Waltron User Manual 3051 Hardness Analyzer ...