Step
Description
Key Points
104-001-A.7 Released 06/14/2019
3.4
Using the Arrow buttons select
O2 AUTO
. Press ENTER button.
In the next screen select
O2 LOW
.
Press ENTER button.
Select
000000 ppb
. Press ENTER
button.
3.5
Calibration will begin. Wait
approximately 2 minutes. Do not
disturb the analyzer or the
calibration equipment.