background image

 

Step 

 

Description 

Key Points 

 

 

104-001-A.7 Released 06/14/2019

 

3.4 

 

Using the Arrow buttons select  

O2 AUTO

.  Press ENTER button. 

 
In the next screen select 

O2 LOW

.  

Press ENTER button. 
 
Select 

000000 ppb

.  Press ENTER 

button. 

 

3.5 

 

Calibration will begin.  Wait 
approximately 2 minutes.  Do not 
disturb the analyzer or the 
calibration equipment. 

 

Содержание 9065

Страница 1: ...104 001 A 7 Released 06 14 2019 9065 Luminescent Dissolved Oxygen Analyzer Calibration Kit Quick Reference Manual...

Страница 2: ...fice 908 534 5100 x120 Waltron s technical expertise and extensive experience provides personalized solutions to the water quality industry It is Waltron s commitment to provide the customer with time...

Страница 3: ...n K6000 101 A calibration beaker is provided with each 9065 analyzer order Contact Waltron for additional beakers p n K5000 039 1 3 Calibration is Due Determine if the Dissolved Oxygen Analyzer requi...

Страница 4: ...flow extra calibration gas or clean dry air over the sensor to expedite drying DO NOT wipe dry the black sensor spot Touching the sensor may damage the delicate coating 1 8 Store the calibration kit...

Страница 5: ...itting A calibration beaker is included with the initial analyzer order Ensure the provided O Ring is installed on the fitting to provide a seal against leaks between the fitting and the beaker 2 2 In...

Страница 6: ...ey Points 104 001 A 7 Released 06 14 2019 2 4 Remove the protective cap from the provided nitrogen tank 2 5 Install the gas regulator on nitrogen gas tank 2 6 Install one end of the provided tubing on...

Страница 7: ...vided tubing on to the barb fitting on the calibration beaker 2 8 Open the regulator all the way to allow flow of Nitrogen gas The provided fixed flow regulator will supply the correct flow rate 2 9 W...

Страница 8: ...d sensor in a dual stream analyzer Press ENTER button Go to O2 SENSOR 1 or 2 CALIBRATION Menu If you have a dual stream analyzer you may select Sensor 1 or Sensor 2 3 3 For oxygen free calibration gas...

Страница 9: ...4 Using the Arrow buttons select O2 AUTO Press ENTER button In the next screen select O2 LOW Press ENTER button Select 000000 ppb Press ENTER button 3 5 Calibration will begin Wait approximately 2 min...

Страница 10: ...Contact Waltron technical support if you are still unsuccessful 3 7 Close the valve on the gas regulator to preserve your calibration gas 3 7 Use the arrow buttons to select STORE CALIBRATION Press EN...

Страница 11: ...Replace plastic cap and store in a clean safe location in the provided kit box Remove the calibration beaker and store in a clean safe location for the next calibration Reinstall the flow cell and the...

Отзывы: