Step
Description
Key Points
104-001-A.7 Released 06/14/2019
3.8
After calibration is successfully
stored,
cycle the power
to reboot
the analyzer by disconnecting and
then reconnecting the power cord
(connector at the bottom of the
transmitter which is furthest to
the left).
3.9
Make sure the valve on the gas regulator is closed.
Remove the gas
regulator
from the gas cylinder. Replace plastic cap and store in a
clean safe location in the provided kit box.
Remove the calibration beaker
and store in a clean safe location for
the next calibration.
Reinstall the flow cell and the inlet and outlet connections. Restore
the sample flow to the analyzer.
Disconnect and store the
calibration kit for the next usage.
Reconnect the analyzer flow cell
and restore sample flow.
The nitrogen gas cylinder can
last up to 5-7 calibrations or
until the expiration date printed
on the cylinder, whichever is
sooner.
3.10
Press
START
button to put the
analyzer back online.