background image

 

Step 

 

Description 

Key Points 

 

 

104-001-A.7 Released 06/14/2019

 

3.8 

 

After calibration is successfully 
stored, 

cycle the power

 to reboot 

the analyzer by disconnecting and 
then reconnecting the power cord 
(connector at the bottom of the 
transmitter which is furthest to 
the left). 

 

3.9 

 
Make sure the valve on the gas regulator is closed. 

Remove the gas 

regulator

 from the gas cylinder.  Replace plastic cap and store in a 

clean safe location in the provided kit box.   
 

Remove the calibration beaker

 and store in a clean safe location for 

the next calibration.   
 

Reinstall the flow cell and the inlet and outlet connections.  Restore  
the sample flow to the analyzer. 
 

Disconnect and store the 
calibration kit for the next usage.   
 
Reconnect the analyzer flow cell 
and restore sample flow.   

 

The nitrogen gas cylinder can 
last up to 5-7 calibrations or 
until the expiration date printed 
on the cylinder, whichever is 
sooner. 

3.10 

 

Press 

START

 button to put the 

analyzer back online. 

 

 

Содержание 9065

Страница 1: ...104 001 A 7 Released 06 14 2019 9065 Luminescent Dissolved Oxygen Analyzer Calibration Kit Quick Reference Manual...

Страница 2: ...fice 908 534 5100 x120 Waltron s technical expertise and extensive experience provides personalized solutions to the water quality industry It is Waltron s commitment to provide the customer with time...

Страница 3: ...n K6000 101 A calibration beaker is provided with each 9065 analyzer order Contact Waltron for additional beakers p n K5000 039 1 3 Calibration is Due Determine if the Dissolved Oxygen Analyzer requi...

Страница 4: ...flow extra calibration gas or clean dry air over the sensor to expedite drying DO NOT wipe dry the black sensor spot Touching the sensor may damage the delicate coating 1 8 Store the calibration kit...

Страница 5: ...itting A calibration beaker is included with the initial analyzer order Ensure the provided O Ring is installed on the fitting to provide a seal against leaks between the fitting and the beaker 2 2 In...

Страница 6: ...ey Points 104 001 A 7 Released 06 14 2019 2 4 Remove the protective cap from the provided nitrogen tank 2 5 Install the gas regulator on nitrogen gas tank 2 6 Install one end of the provided tubing on...

Страница 7: ...vided tubing on to the barb fitting on the calibration beaker 2 8 Open the regulator all the way to allow flow of Nitrogen gas The provided fixed flow regulator will supply the correct flow rate 2 9 W...

Страница 8: ...d sensor in a dual stream analyzer Press ENTER button Go to O2 SENSOR 1 or 2 CALIBRATION Menu If you have a dual stream analyzer you may select Sensor 1 or Sensor 2 3 3 For oxygen free calibration gas...

Страница 9: ...4 Using the Arrow buttons select O2 AUTO Press ENTER button In the next screen select O2 LOW Press ENTER button Select 000000 ppb Press ENTER button 3 5 Calibration will begin Wait approximately 2 min...

Страница 10: ...Contact Waltron technical support if you are still unsuccessful 3 7 Close the valve on the gas regulator to preserve your calibration gas 3 7 Use the arrow buttons to select STORE CALIBRATION Press EN...

Страница 11: ...Replace plastic cap and store in a clean safe location in the provided kit box Remove the calibration beaker and store in a clean safe location for the next calibration Reinstall the flow cell and the...

Отзывы: