6. BUFFER SOLUTION REFERENCE TABLE
While using standard buffer solutions for calibration, use only
Buffer solution with the following temperature coefficient for
accurate calibration:-
Temperature
(°C)
Buffer 7
(pH)
Buffer 4
(pH)
Buffer 10
(pH)
0
7.11
4.00
10.32
5
7.08
4.00
10.25
10
7.06
4.00
10.18
15
7.03
4.00
10.12
20
7.01
4.00
10.06
25
7.00
4.01
10.01
30
6.98
4.02
9.97
35
6.98
4.02
9.93
40
6.97
4.03
9.89
45
6.97
4.04
9.86
50
6.97
4.06
9.83
55
6.97
4.07
9.80
60
6.98
4.10
9.78
7.
OPTIONAL ACCESSORIES
Description
Order Code
pH Electrode
-
PE-03
ORP Electrode
-
ORP106
Temperature probe
-
TP-200
Buffer Solution pH 7.00
-
7010
Buffer Solution pH 4.01
-
4010
Buffer Solution pH 10.01
-
1010