![Walker Electronics Q105 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/walker-electronics/q105/q105_operators-handbook-manual_870416022.webp)
Page 21
the validation report printout explained...
When a validation receipt is printed it may look similar to the one below
Whilst the above is mostly self-explanatory...
Duration 360 seconds of 360. This means that the timer was set for 6
minutes and completed its full set cycle.
Mean power. This indicates the average power supplied to the transducer
through the cycle. A figure of below 100 watts on 100% power
may
indicate a
problem with the unit. This could just be due to fluid level or
contamination. Please call Walker Electronics Ltd
for further guidance.
Energy 15570 of 18000 joules. The machine adjusts the total amount of
cleaning power that has been applied during a complete cycle. If this total
power for the time (known as joules) is below the expected level the
machine adjusts the cleaning time and extends the cycle accordingly.