9
WDIS410 or WDS100 or WDS600 or W900
The sensor is provided with a 2-twisted pair, shielded, 24 AWG, 35 pF/foot
capacitance cable. The wiring to the controller is as follows:
Shield
Drain:
Earth Ground
GRN:
IN+
WHT:
IN-
RED:
+5 V
BLK:
- 5 V
If the required cable length exceeds the 6 meters (20 feet) that is supplied,
wire the housing to a 190851 terminal box, then use a p/n 100084 cable to
reach the instrument. The maximum cable length is 30.5 meters (100 feet).
WDIS410
Power Supply
(115 VAC or 230 VAC)
Contact Closure:
Polarity not critical
Interlock Function
+5V
T-
T+
GROUND
STUD
L1 L2/N
G
R
N
1
20
V
G
R
N
/Y
E
L 2
40
V
WHT 120V
BLU 240V
B
L
K
1
2
0
V
B
R
N
2
4
0
V
IN
+
IN
-
+
5
V
-5
V
-5V
IN+ IN-
IN-
DIG IN 2 DIG IN 3
IN+ IN-
DIG IN 1
IN+
IN-
IN+
DIG IN 4
IN+ IN-
DIG IN 5
IN+ IN-
IN
-
D
IG
IN
2
D
IG
IN
3
IN
+
IN
-
D
IG
IN
1
IN
+
IN
-
IN
+
D
IG
IN
4
IN
+
IN
-
D
IG
IN
5
IN
+
IN
-
L1
L2/N
L2
L2
L2
L2
L2
L2
BLEED
N.C.
N.O.
N.C.
BOI 1
N.O.
N.C.
FEED
N.C.
BIO 2
N.O.
N.C.
N.O.
N.O.
ALARM
N.C.
N.O.
G
R
E
E
N
W
H
IT
E
R
E
D
BL
AC
K
Disinfection
Sensor