8
Note: 1 U.S. Gallon = 231 U.S cubic
inches
1 Liter = 61.03 U.S. cubic inches
Volume of Cooling Coil: 0.018 Gallons
0.068
Liters
Volume of Cooling Plate: 0.023 Gallons
0.088
Liters
Volume of 3/8" O.D. x 1/4" I.D. (0.59 in3/ft):
0.00255 Gallons/linear ft
0.00965
Liters/linear
ft
Volume of the system =
17.7 in3
= 0.0765 gallons
231
in3
/
gallon
Maximum lagtime
=
0.60 minutes (previously calculated)
So, the minimum pump flow rate
=
0.0765
gallons
0.60
minutes
=
0.127
gal/min
(483
mL/min)
Caution: The calculated pump flow rate is the minimum required to obtain the
desired control band, however, if the flow rate increases over the recommended rate
of 500 mL/min (approx. 0.13 gal/min) the rate of cooling will decrease. This may be
compensated for by re-evaluating the system criteria: length / desired control band
or to double up on the cooling plate/coil.
Consult factory with any further installation questions.
3.5
Control Module Installation
Once the enclosure is mounted, the output pumps may be located at any distance
from the controller. The sensor may be placed up to 80 feet from the controller.
Shielded cable is preferable. Always route AC voltage wiring in conduit that is
separated a minimum of 6 inches away from low voltage DC signal lines (such as
the sensor signal).