12.0 CALIBRATION SERVICE
Recalibration service can be provided at a nominal cost, if desired. Special thermal calibration
traceable to NIST is available. Standard temperature points are: 0°C, 94°C, and 250°C. Custom
points are available upon request.
TOLERANCE (PROBE INTERCHANGEABILITY)
°F
0
-20
-40
-60
-80
-100
-120
-140
-160
-180
-200
0
93.03
88.65
84.26
79.85
75.42
70.98
66.52
64.04
57.54
°F
0
20
40
60
80
100
120
140
160
180
200
0
93.03
97.39
101.74
106.07
110.38
114.68
118.97
123.24
127.50
131.74
135.97
200
135.97
140.18
144.38
148.57
152.74
156.90
161.04
165.17
169.28
173.28
177.47
400
177.47
181.54
185.60
189.64
193.67
197.69
201.69
205.68
209.65
213.61
217.55
600
217.55
221.49
225.40
229.31
233.20
237.07
240.93
244.78
248.61
252.43
265.23
800
256.23
260.02
263.80
267.56
271.31
275.04
278.76
282.46
286.15
289.83
293.49
TOLERANCE (PROBE INTERCHANGEABILITY)
Temp
°C
-420
-400
-350
-300
-200
-100
0
32
100
200
300
400
500
600
700
800
900
Tolerance
±°C
10.4
8.3
3.3
1.9
1.5
1.1
0.7
0.6
0.7
0.9
1.5
2.0
2.7
3.3
3.9
4.5
5.4
±:
1.0
0.9
0.7
0.5
0.4
0.3
0.2
0.1
0.2
0.2
0.3
0.4
0.6
0.7
0.8
0.9
1.0
TOLERANCE (PROBE INTERCHANGEABILITY)
Temp °C
-250
-220
-200
-100
0
100
200
300
400
500
Tolerance
±°C
5.8
1.8
1.2
0.7
0.3
0.5
1.1
1.7
2.3
3.0
±:
1.0
0.7
0.5
0.3
0.1
0.2
0.4
0.6
0.8
1.0