Malfunction
Possible cause
Remedy
Link
Instrument table
not in moveable
in working
positon
Ball locking jams or is
incorrectly adjusted
Contact distributor
page 21
Table bearing damaged
Contact distributor
page 21
Instrument table
not fixed in
working position
Holding magnet
damaged
Contact distributor
page 21
Switching relay damaged
Contact distributor
page 25
End position switch
damaged
Contact distributor
page 21
Electro-
magnetic
See below
(Table E- interference)
Increase distance,
Potential equalization
Eliminating electromagnetic interference
Service
Band (MHz)
Test frequency
(MHz)
Max. power
P(W)
Distance
d (m)
Different services
(TETRA 400)
380-390
385
1,8
1
FRS-460, GMRS 460
430-470
450
2
1
LTE Band 13/17
704-707
710/745/780
0,2
1
GSM800/900
LTE Band 5
Mobile phone CT1+, CT2, CT3
800-960
810/870/930
2
1
GSM1800/1900
DECT (mobile phone)
LTE Band 1/3/4/25
UMTS
1700-1990
1720/1845/1970
2
1
Bluetooth, WLAN 802.11b/g/n
LTE Band 7
RFID 2450 (active, passive)
2400-2570
2450
2
1
WLAN 802.11a/n
5100-5800
5240/5500/5785
2
1
Caution!
Portable HF telecommunication devices may not be used at a distance of less than 1 meter from the
device including the cables.