15
Wagner & Guder Medical GmbH
issue May 2021
Direction for use, cubeLITE
Cleaning
For cleaning the varnished parts and the seat upholstery of cubeLITE
please use a wet cloth. Into the water you can add a mild cleaning agent
(if necessary „scouring milk“). Obstinate dirt at varnished surfaces can be
removed with the help of cleaning solvent or rectified spirits.
In case of cleaning the installed examination devices, please follow the
hints and comments of each device producers or ask authorized
specialist staff.
Maintenance/ Technical service
The examination unit cubeLITE and its accessories are low-
maintenance when proper used. Nevertheless periodic maintenance
works, which serve the units safety, have to be carried out either self-
supporting or by our service partners:
1
Check the proper fastening of the examination devices regularly.
2
Check the safety advices affixed at the supplied products.
3
Check the safety inscriptions at your examination unit. When in-
scription elements are missing or damaged contrary to the user
manual please contact our service technician.
4
Check the integrity of housing an isolations
5
Check the operational reliability of safety-relevant circuits (switch-
off panel and current limitations)
6
Regular measuring (annual) of substitute leakage current
7
Complete check of examination unit’s and accessory's operational
reliability
In case of technical problems or decline of handling please consult your
dealer or our service center.