SENSOR FAILURE REQUIRES REPAIR
This indicates that the light source is functioning, but the receiver is not receiving the output.
A dirty lens on either the transmitter and/or the receiver side. Malfunction in the sensor head assembly or
cable.
Action:
Clean the lenses as described in “Maintenance” section. If this does not correct the problem, call WAGER
Technical Services for assistance.
RECHARGE BATTERY
The battery is fully discharged. This message is displayed for a few seconds before the unit turns itself off.
Action:
Recharge the battery or run the unit using the AC power adapter.
BATTERY LOW RECHARGE SOON
The battery is getting low and needs recharging.
Action:
Recharge the battery or run the unit using the AC power adapter. (The unit will continue to function normally
after displaying this message.)
MEMORY FULL DELETE TESTS FIRST
The unit saved 100 tests and has reached full storage capacity. Either you have saved 100 tests, or the unit
has experienced some electrical shock that has reset the number of saved tests. The unit immediately goes
into the Print and Delete Test Series mode after displaying this message.
Action:
Use the ZERO to delete test series.
PRINTER NOT ATTACHED
The unit could not find a printer attached to the serial port.
Action:
Make sure both ends of the printer cable are firmly attached. Also make sure the printer is configured for serial
operation (see documentation in Appendix F).