Wachendorff PH5100 Скачать руководство пользователя страница 5

Operating instruction BA_074_2020-01_10.00  Universal panel indicator PH5100

Page 5/15

  

Operating manual 

 

- 5 -    

BA_074e, Rev.1C5 

The alarms can assume the functions described on 

Table 02

off

 

Alarm off. 

        



 

Alarm of the Absolute Minimum 

Value. It triggers when the value 

of the 

PV

 is 

below

 the value 

defined by the alarm 

Setpoint 

(SPA1 or SPA2)

SPA1

PV

   







 

Alarm of the Absolute Maximum 

Value. It triggers when the value 

of the 

PV 

is 

above

 the value 

defined by the alarm 

Setpoint

SPA1

PV

 

 

 

 





 

Alarm of the Differential Value. In this function the parameters “

SPA1

” and “

SPA2

” 

represent errors (difference) between the PV and one reference value (

AlrF

). 

ALrF

PV

ALrF+SPA1

ALrF–SPA1

ALrF

PV

ALrF –SPA1

ALrF + SPA1 

alarme

SPA1 positive 

SPA1 negative 







  

Alarm of the Minimum Differential Value. It triggers when the value of the PV is 

below

the point defined by:

ALrF-SPA1

 (using alarm 1 as an example). 

ALrF

PV

ALrF – SPA1

ALrF

PV

ALrF – SPA1

SPA1 positive 

SPA1 negative 







  

Alarm of the Maximum Differential Value. It triggers when the value of the PV is 

above

the point defined by: 

ALrF+SPA1

 (using alarm 1 as an example). 

ALrF

PV

ALrF + SPA1

ALrF

PV

ALrF + SPA1

SPA1 positiv 

SPA1 negativ 







 

 

 

 

Alarm of the Absolute Minimum Value. It triggers when the value of the totalizer (TOT) is 

below

 the value defined by the alarm 

Setpoint. 







 

 

 

 

Alarm of the Absolute Maximum Value. It triggers when the value of the totalizer (TOT) 

is 

above

 the value defined by the alarm 

Setpoint.







 

 

 

 

Alarm at overflow of totalizer. 







 

 

 

 

Alarms of the Sensor Break (Sensor 

Break Alarm

). It is activated when the Input 

presents problems such as interrupted sensor, bad connection, etc. 

Table 02

: Alarm functions 

Note:

 The figures are also valid for Alarm 2 (SPA2). 

Содержание PH5100

Страница 1: ...H Im Gewerbepark 13 27798 Hude Germany Hotline 49 4408 804 444 Fax 49 4408 804 100 e mail sales labom com www labom com Operating instruction BA_074_2020 01_10 00 Page 1 3 BA_074e Rev 1C5 info labom c...

Страница 2: ...on scale factor t ScF 6 4 9 Batch totalizing function 6 4 10 F1 and F2 function keys 6 4 11 Over flow alarm 7 4 12 Minimum and maximum 7 4 13 24 VDC auxiliary voltage source 7 4 14 Interface 7 5 Insta...

Страница 3: ...5100 are not meant to be used as sole safety means to prevent dangerous situations on machinery and installations Machinery and installations must be so designed that fault conditions can not lead to...

Страница 4: ...t Type CODE Measurement range Thermo couple J Range 110 C to 950 C 166 F to 1742 F Thermo couple K Range 150 C to 1370 C 238 F to 2498 F Thermo couple T Range 160 C to 400 C 256 F to 752 F Thermo coup...

Страница 5: ...he Minimum Differential Value It triggers when the value of the PV is below the point defined by ALrF SPA1 using alarm 1 as an example ALrF PV ALrF SPA1 ALrF PV ALrF SPA1 SPA1 positive SPA1 negative A...

Страница 6: ...stored in a non volatile internal memory and can also be presented at the indicator display by choosing TOTAL screen which is identified by the symbol TOT This symbol is always presented when the TOTA...

Страница 7: ...es up to 99999 99999 or 9999 99999 The flags HIGH and LOW identify if the part being displayed is the SP most significant part HIGH or the least significant part LOW The totalization alarm SP is defin...

Страница 8: ...pplication Figure 04 Auxiliary 24 Vdc usage example 5 3 Recommendations for the installation To minimize the pick up of electrical noise the low voltage DC connections and the sensor input wiring shou...

Страница 9: ...nsisting of assigning values to the parameters according to the desired behavior The user shall understand the importance of each parameter and determine a valid condition or a valid value for each on...

Страница 10: ...imum differential Alarm Referenz Totalization alarm SP decimal point SetPoint Alarm Alarm SP Value that defines the point of activation of the alarm outputs For the alarms programmed with the function...

Страница 11: ...it It defines the upper value of the indication range when the input types of 0 20 mA 4 20 mA 0 50 mV 0 5 V and 0 10 V are used avto r Defines the function to be executed by F1 and F2 keys No associat...

Страница 12: ...ferent from zero Protection Sets up the Level of Protection See Table 03 page 13 Frequency Frequency of the local electrical network Serial number 4 most significant digits Serial number 4 least signi...

Страница 13: ...nded for allowing the user to define a new password in the event of it being forgotten The Master Password doesn t grant access to all parameters only to the Password Change parameter PAS After defini...

Страница 14: ...0 2 of the span Input impedance Pt100 T C 0 to 50 mV 10 M 0 to 5 V 0 to10 V 500 k 4 to 20 mA 100 Measuring of the Pt100 3 wire type 0 00385 with compensation of the cable length max 50 meters excitat...

Страница 15: ...the device described therein 12 Disclaimer All technical content within this document can be modified without notice The content of this document is content of a recurring audit In losses due fire ea...

Отзывы: