28
Service
Renewing the filter pack
> Do not touch the vents of the suction filter (risk of damage)!
> Use the W&H Assistina for maintenance only when the suction filter with a fully serviceable filter pack is fitted.
> Renew the filter pack in the suction filter after approx. 7,000 maintenance cycles (corresponds to the consumption of a 500 ml
bottle of W&H service oil).
Pull out the suction filter
horizontally towards the rear.
Press out the filter pack using
a suitable implement.
Ensure that it is placed in the
correct waste disposal.
Take the filter pack out
of its packaging.
Press the filter pack on the
frame in the filter case.
Refit the suction filter so that
it is leak-tight.