17
10. Maintenance
> First carry out the external disinfection and cleaning.
> Carry out daily maintenance midday and evening prior to sterilization.
> Use the adaptor REF 02679000 for maintenance of transmission instruments in those cases where the manufacturer
has allowed for removal of the treatment heads.
With sterilization after every patient
> Press the program button once
With maintenance twice a day
> Press the program button twice on each occasion
An accurate dosage of W&H cleaning fluid and W&H service oil is provided for these maintenance processes.
Additional pressing of the program button can cause too much cleaning fluid to enter the spray air and coolant channels!