EX7000 Series Web Page Operation
F
IGURE
4-10:
U
PGRADE
W
EB
P
AGE
Perform the following steps to conduct a firmware upgrade:
1)
Perform a
Reboot
or a power cycle.
2)
Connect to the EX7000 via the embedded web page.
3)
Click on the
Upgrade
link.
4)
Click on the
Browse
button and select the firmware image file to be uploaded to the
instrument.
5)
Click the
Submit
button to initiate the upgrade process.
The upgrade process takes approximately 5 minutes to complete and will be followed by a prompt
to reboot the instrument. Once the reboot is complete, reconnect and confirm on the
Index
page
that the firmware revision level has been properly updated.
R
ESET
This action page is used to return the EX7000 to a known communication state. Pressing the Reset
button will cause the EX7000 to terminate all active sessions, leaving only the current session
viable.
R
EBOOT
This action page is used to perform a complete instrument reboot, equivalent to that which occurs
when the instrument is power cycled. It is most commonly used to accept changes that are made to
the network configuration or time configuration settings. In addition, it is suggested that a reboot
be performed before conducting a firmware upgrade.