EX7000 Series Basic Operation 

21 

When enabled as an output, each channel also has the ability to generate a 1 µs pulse upon 
command. An example application of this pulse is to use the EX7000 to externally trigger another 
piece of test equipment. The specific operation of the pulse depends on the static level 
programmed for that channel. When a channel is programmed with a static level of high, the pulse 
will be low-going. When a channel is programmed with a static level of low, the pulse will be 
high-going. Each pulse generation requires a separate command. 

The electrical specifications for the digital I/O port are provided in Table 3-1. Particular note 
should be given to the 

V

INPUT

 specification of -0.5 V to 5.5 V. Exceeding that value with an 

external voltage source, even through a resistance, could permanently damage the EX7000. 

Characteristic 

Value 

V

INPUT 

-0.5 V to 5.5 V 

V

IH

 

2 V min 

V

IL 

0.8 V max 

V

OH

 (

I

OH

 = -32 mA) 

2 V min 

V

OL

 (

I

OL

 = 64 mA) 

0.55 V max 

 
 

T

ABLE 

3-1:

 

D

IGITAL 

I/O

 

P

ORT 

E

LECTRICAL 

S

PECIFICATIONS

 

General Purpose I/O 

The EX72SF features an additional 32-bit general purpose TTL I/O port. This is in additional to 
the internal 72-channel relay drivers and the 8-bit digital I/O described above. The following are 
the pin assignments for the rear-mounted 37-pin D-sub connector: 

Pin 

Signal 

Pin 

Signal 

1 GND 

20 

PIO-PORT0 

2 PIO-PORT1 

21 

PIO-PORT2 

3 PIO-PORT3 

22 

PIO-PORT4 

4 PIO-PORT5 

23 

PIO-PORT6 

5 PIO-PORT7 

24 

PIO-PORT8 

6 PIO-PORT9 

25 

PIO-PORT10 

7 PIO-PORT11 

26 

PIO-PORT12 

8 PIO-PORT13 

27 

PIO-PORT14 

9 PIO-PORT15 

28 

GND 

10 +5 

29 GND 

11 PIO-PORT16 

30 PIO-PORT17 

12 PIO-PORT18 

31 PIO-PORT19 

13 PIO-PORT20 

32 PIO-PORT21 

14 PIO-PORT22 

33 PIO-PORT23 

15 PIO-PORT24 

34 PIO-PORT25 

16 PIO-PORT26 

35 PIO-PORT27 

17 PIO-PORT28 

36 PIO-PORT29 

18 PIO-PORT30 

37 PIO-PORT31 

19 GND 

 

 

 

 

A current limited supply of +5 V is provided on pin 10. An internal, resettable fuse is provided to 
protect 200 mA against excessive current flow and short circuits. 

Содержание EX7000 SERIES

Страница 1: ...EX7000 SERIES MICROWAVE SWITCH ASSEMBLIES WITH LXI COMPLIANT LAN INTERFACE USER S MANUAL P N 82 0121 000 Release December 14 2007 VXI Technology Inc 2031 Main Street Irvine CA 92614 6509 949 955 1894...

Страница 2: ...VXI Technology Inc 2 EX7000 Series Preface...

Страница 3: ...mon Features 10 Interactive Control 11 Relay Drive Control 12 Interfacing with Other Components 13 Features 13 Relay Odometer 13 Exclude Lists 13 Scan Lists 13 Configuration Table 13 Programmable MBB...

Страница 4: ...oft Front Panel 28 Network Configuration 29 Time Configuration 31 Device Identify 33 Upgrade 33 Reset 34 Reboot 34 SECTION 5 35 THEORY OF OPERATION 35 Introduction 35 System Overview 35 Open Collector...

Страница 5: ...s for products returned to VTI from another country VTI Microwave warrants that its software and firmware designated by VTI Microwave for use with a product will execute its programming when properly...

Страница 6: ...equirements of the Low Voltage Directive 73 23 EEC and the EMC Directive 89 366 EEC inclusive 93 68 EEC and carries the CE mark accordingly The product has been designed and manufactured according to...

Страница 7: ...ce with EN 50419 Marking of electrical and electronic equipment in accordance with Article 11 2 of Directive 2002 96 EC WEEE End of life product can be returned to VTI by obtaining an RMA number Fees...

Страница 8: ...A 92614 6509 Phone 949 955 1894 Fax 949 955 3041 VXI Technology Cleveland Instrument Division 5425 Warner Road Suite 13 Valley View OH 44125 Phone 216 447 8950 Fax 216 447 8951 VXI Technology Lake Ste...

Страница 9: ...plus any additional active or passive components that may be required Every member of the EX7000 series comes with an LXI compliant LAN interface EX71HD High Density 26 GHz Modular Microwave Switchin...

Страница 10: ...allowing a customer to add their own RF and microwave components to the front and rear panels as well as the removable mezzanine that is located at approximately the mid way point between the top and...

Страница 11: ...g which provides a JAVA applet for interactively configuring the box for its specific configuration as well as allowing for the direct control of components A functional list of the actual components...

Страница 12: ...gh current drivers which are divided into twelve sections of six lines Each section consists of six high current drive lines six indicator status bits one reset line four relay ID bits one relay power...

Страница 13: ...r Each relay drive line has an associated odometer which tracks the number of times a relay coil has been actuated This data is written to non volatile memory and can be called from application code o...

Страница 14: ...Flash ROM Single command response time 32 MB 8 MB 3 ms RELAY DRIVER BOARD RDB RELAY POWER 5 V to 48 V Three power pins each connected to four relay driver connections RELAY DRIVERS 72 open drain high...

Страница 15: ...e cord All components should be immediately inspected for damage upon receipt of the unit INSTALLATION LOCATION The EX7000 is designed to be largely insensitive to external electrical magnetic and the...

Страница 16: ...FIGURATION With its default network configuration the EX7000 will attempt to locate a DHCP server If one is found the IP address assigned by the DHCP server will be used Otherwise after a timeout of 2...

Страница 17: ...0D 3F 01 FE FE 169 254 254 254 00 0D 3F 01 FF FE 169 254 254 254 TABLE 2 1 AUTOIP DEFAULT ADDRESS ASSIGNMENT If a static IP address assignment is preferred one can be optionally assigned via the embe...

Страница 18: ...VXI Technology Inc 18 EX7000 Series Preparation for Use...

Страница 19: ...ccess and control of the internal microwave components REAR PANEL VIEW The rear panel of the EX7000 contains an ac power entry module an ac power switch an Ethernet 10 100 Base T connection a general...

Страница 20: ...O port on the rear panel of the instrument This port can be used as a trigger source and as a general purpose output device The digital I O connector is a standard DB 9 with the following pin assignm...

Страница 21: ...UT 0 5 V to 5 5 V VIH 2 V min VIL 0 8 V max VOH IOH 32 mA 2 V min VOL IOL 64 mA 0 55 V max TABLE 3 1 DIGITAL I O PORT ELECTRICAL SPECIFICATIONS General Purpose I O The EX72SF features an additional 32...

Страница 22: ...contains two rear facing alignment pins that fit into holes on the front panel These pins only allow full insertion of the switch into the EX71HD enclosure opening when properly aligned NOTE Before p...

Страница 23: ...are mounted on special carriers for installation into the EX72SF and are delivered this way if ordered separately from the factory The SP4T and SP6T switches mount directly in the lower openings and...

Страница 24: ...nsert the ribbon cable connector into the mating connector on the switch module Align the ribbon cable connector tab with the release tab on the switch s connector Press the ribbon cable connector fir...

Страница 25: ...ng Model Manufacturer Serial Number Description LXI Class LXI Version Hostname MAC Address IP Address Netmask Firmware version This page is accessible from any other instrument page by clicking on the...

Страница 26: ...screen is done with the Tab key The Enter key has the same function as clicking the Submit button and cannot be used for navigation FIGURE 4 2 EX7000 MAIN WEB PAGE Password Protection Once the IP add...

Страница 27: ...security should consider using private dedicated subnets firewalls etc to limit access to the instrument on the network VXI Technology Logo The VXI Technology logo that appears on the upper left of a...

Страница 28: ...T PANEL An example of the EX7000 Soft Front Panel page can be seen in Figure 4 4 By accessing this page the various switch modules attenuators and other devices connected to the EX7000 can be viewed i...

Страница 29: ...he second to last and last octets of the device s MAC address However X will be set to 1 if it is 0 in the MAC address and to 254 if it is 255 in the MAC address for conformance with the AutoIP specif...

Страница 30: ...he DHCP server would be assigned to the EX7000 at power up initialization without having to manually set it on the EX7000 The DHCP server configuration provides a centralized controlled database of as...

Страница 31: ...he instrument cannot reach the Internet Manual time entry is not affected by the Time Zone control and does not require an instrument reboot to be activated However manually specified time is volatile...

Страница 32: ...VXI Technology Inc 32 EX7000 Series Web Page Operation FIGURE 4 7 TIME CONFIGURATION WEB PAGE SNTP FIGURE 4 8 TIME CONFIGURATION WEB PAGE PTP...

Страница 33: ...B PAGE UPGRADE This action page is used to upgrade the embedded firmware of the EX7000 Prior to initiating the firmware upgrade process a new uncompressed firmware image must be obtained from VXI Tech...

Страница 34: ...to reboot the instrument Once the reboot is complete reconnect and confirm on the Index page that the firmware revision level has been properly updated RESET This action page is used to return the EX...

Страница 35: ...1 FUNCTIONAL BLOCK DIAGRAM SYSTEM OVERVIEW Using the drivers provided with the EX7000 commands are issued from the host PC to the Digital Controller Board DCB The DCB in turn controls the operation o...

Страница 36: ...system OPEN COLLECTOR VS TTL DRIVE LINES The EX7300 and EX7400 relay drive connectors reside on a RDBs that are partitioned into three banks J2 through J5 constitute bank 0 J6 through J10 bank 1 and J...

Страница 37: ...k the necessary current through the coils effectively switching the relay The user can program the relay settling time for each RDB The relay control can be operated in two modes Pulse and Continuous...

Страница 38: ...This value will be used as both pulse width and settling time FIGURE 5 5 PULSE OPERATION MODE NOTE There is no current monitoring capability on the EX7000 It is the user s responsibility to estimate...

Страница 39: ...n Initialize LXI Trigger Bus 20 M MAC address 16 29 MDSN See multicast domain name system multicast domain name system 30 N network configuration resetting See LAN configuration initialize network con...

Отзывы: