H08.0.1X.6C-18_VWR
Specification
Measuring range:
-199.9 ... +199.9 °C
Accuracy:
± 0.1 % of meas. value ± 2 digit (at nom. temperature)
(at range
–70.0 ... +199.9 °C – at low er values refer to c orrection table below )
Sensor:
Pt1000, electrically insulated and mounted in stainless
steel tube of 3 mm dia and approx. 100 mm length,
temperature-resistant up to 200 C
Probe permanently connected to the device.
Core temperature probe: small Teflon handle ~ 75 mm
1 m Teflon cable,
handle and cable temperature-resistant up to 250 C
Resolution
0.1 °C
Display:
approx. 13 mm high, 3½-digit LCD
Operating elements:
switch for on/off, 2 adjusting potentiometer
Nominal temperature:
25 °C
Working condition:
-30 to +45 °C, 0 to 80 %RH (non condensing)
Storage temperature:
-30 to 70 °C (device)
Power supply:
9V-battery type 6F22 (included)
Power consumption
1.4 mA (typ.)
Battery life
~ 200 operating hours
Low battery warning:
"BAT" displayed automatically in case of low battery
Housing:
Impact-resistant ABS-housing, front IP65
Dimensions:
~ 106 x 67 x 30 mm (H x W x D), without probe
Weight:
~ 190 g incl. battery and probe
EM C:
The device corresponds to the essential protection ratings
established in the Regulations of the Council for the A p-
proximation of Legislation for the member countries re-
garding electromagnetic compatibility (2004/108/EG).
Additional error: <1%
Correction table for low temperatures
temperature display
-200 -201.9
-195 -196.7
-190 -191.6
-185 -186.4
-180 -181.2
-175 -176.1
temperature display
-170 . -171.0
-165 -165.9
-160 -160.8
-155 -155.7
-150 -150.6
-145 -145.5
temperature display
-140 -140.4
-135 -135.4
-130 -130.3
-125 -125.3
-120 -120.2
-115 -115.2
temperature display
-110 -110.1
-105 -105.1
-100 -100.1
-90 -90.0
-80 -80.0
-70 -70.0