VWR Collection PCR Workstation
25
Replacing the UV Air Recirculator UV tube
Locate and remove the six screws on the ventilation cover on
the outer back panel of the workstation.
Pull away the ventilation cover from the workstation.
The tube has two prongs on each end that fit into a socket, then
rotate into place.
Carefully hold the tube at the ends and insert the tube into the
socket. Twist the tube 90° until the tube locks into place.
Replace the cover back into position and reattach the six
screws.
We recommend you change the UV tubes after 1000 – 1500
hours of operation.
Replacing the filter of the UV Air Recirculator
The UV Air Recirculator fan filter is located on the inner top of
the PCR Workstation
in front of the light tubes (see picture to the
right). To change the filter, take off the plastic cover with the filter
attached to it and replace it with a new filter. Push the plastic
cover back into its location.
Replacing the fuse
The fuse is located at the back-right side of the system. To remove
the fuse, use a flathead screwdriver to open the lid and pull out the
fuse.
Insert the new fuse and using the screwdriver, turn the fuse clockwise
to lock.