Operation
OX 4000 H
38
ba75853e04
06/2012
5.6
Reset
You can reset (initialize) the sensor settings and sensor-independent
settings separately from each other.
5.6.1
Resetting the measurement settings
The following measuring parameters are reset to the delivery status:
All system settings are retained.
Note
The measuring system is not calibrated after a reset. Before measur-
ing, recalibrate the meter.
Resetting the
measurement settings
Measuring mode
D.O. concentration
Relative slope
1.00
Salinity
0.0
Salinity correction
Off
1
Press
<ON/OFF
_
>
to open the menu for the reset of the mea-
surement settings.
Ini.C
is displayed.
2
Press
<
><
>
to display
no
or
YES
.
YES
: Reset the measurement settings.
no
: Retain the measurement settings.
3
Confirm with
<OK>
.
The menu is finished.
The meter switches to the measuring mode.
o
n
C
i
n
I
Cal