![VWR OX 4000 H Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/vwr/ox-4000-h/ox-4000-h_operating-manual_1057580031.webp)
OX 4000 H
Operation
31
ba75853e04
06/2012
5.4.1
Options for data transmission
Via the USB interface you can transmit data to a PC.
The following table shows which data are transmitted to the interface in
which way:
5.4.2
Downloading stored measurement datasets
5.4.3
Downloading calibration data
Data
Control
Operation / description
Current measured
values
Manual
With
<OK>
.
Simultaneously with every
manual storage process
(see section 5.3.1).
Stored measured
values
Manual
All datasets
(see section 5.4.2)
Calibration records
Manual
Without display indication
(see section 5.4.3).
During the display indica-
tion with
<CAL
_
>
(see section 5.3.4).
Automatic
At the end of a calibration
procedure.
1
Open the storage menu with
<RCL>
.
2
If necessary, scroll with
<RCL>
until
Sto Prt
is displayed.
3
Press
<OK>
to download the measurement data to the inter-
face.
1
Open the storage menu with
<RCL>
.
2
If necessary, scroll with
<RCL>
until
CAL Prt
is displayed.
3
Press
<OK>
to download the calibration data to the interface.