- 43 -
-
"
Download calibration data
" enables downloading the calibration data
from the meter. Choose the measuring function in the meter to collect its
report.
In this option choose:
„Name“
- the meter's name will be given to each report.
At least one of the other following options
needs to be selected with this one;
"
Factory data
"
- downloads the meter‘s data: name, serial
number calibration date;
"
Last calibration
"
- downloads only the last calibration in a chosen
measuring function;
"
History of calibration
"- downloads all the stored calibrations in a
chosen measurement function. The first
downloaded calibration will be the most recent
one.
"
Signatures
"
- space for a handwritten signature.
The data will be downloaded after choosing an option and pressing the
"START"
button.
Note:
before downloading the last calibration or the history of
calibrations, mark the pH or the conductivity value with the
green frame.