- 24 -
12.2. Measurement with manual temperature compensation
Disconnecting the temperature probe from the meter switches the meter to
the manual temperature compensation mode. Next to the entered
(not measured) temperature value, instead of the symbol, the symbol is
displayed (Pic. 8). The procedure of measurement with manual temperature
compensation is similar to that of measurement with ATC. The difference is in
entering the solution temperature value (description – point 15.5). This value
is displayed in place of the measured temperature and is used for
compensation.
Manual compensation may be used in stable conditions, i.e. during laboratory
pH measurements, especially with use of thermostat, or in case of the
temperature probe breakage.
In case of measurement with manual temperature compensation:
- turn the meter on with the
button;
- insert the pH electrode into the vessel with the measured solution; if the
electrode is not calibrated or has already been in use for a long period of
time, perform a calibration (chapter 10). During measurements in a vessel
don’t touch its bottom and walls with the electrode. It is advisable to use an
electrode stand;
- measure the temperature of the solution with use of a laboratory
thermometer and enter the value according to the point 15.5;
- wait till the value stabilises and read the result.
Pic. 8
User‘s manual for
PH-2500L
pH meter
v1.0x
Содержание 662-2380
Страница 1: ...Instruction manual VWR pH METER PH 2500L EU cat no 662 2380 Version 1 1 Issue 12 2021 ...
Страница 5: ... 1 I INTRODUCTION ...
Страница 14: ... 10 II pH MEASUREMENT User s manual for PH 2500L pH meter v1 0x ...
Страница 29: ... 25 III REDOX POTENTIAL AND TEMPERATURE MEASUREMENT ...
Страница 36: ... 32 IV OPTIONS User s manual for PH 2500L pH meter v1 0x ...
Страница 39: ... 35 Pic 14 ...