![VWR 662-2351 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/vwr/662-2351/662-2351_instruction-manual_1057557034.webp)
- 30 -
VWR Collection
PH-130
pH-meter instruction manual
v1.0x
20. TECHNICAL DATA
pH MEASUREMENT:
range
resolution
accuracy
(±1 digit)
0.00
14.00 pH
0.01 pH
±0.01 pH
Input impedance:
>1012
Temperature compensation:
manual/automatic
Compensation range:
-5.0 ÷ 110.0 °C
pH electrode calibration:
automatic,
in 1
3 points
pH electrode calibration range:
Offset:
±0.7 pH
Efficiency:
85 % ÷ 105 %
Thermal stability of zero:
0.001 pH/ °C
mV MEASUREMENT:
range
resolution
accuracy
(±1 digit)
-2000.0
2000.0 mV
1 mV
±1 mV
Input impedance:
>1012
Relative measurement range:
±3999.9mV
TEMPERATURE MEASUREMENT:
range*
resolution
accuracy**
(±1 digit)
- 50.0 ÷ 200.0 °C
0.1 °C
0.1 °C
* The accessories connected to the meter may have narrower working temperature
range.
** Accuracy given for the meter. The final accuracy depends on the type of Pt-1000
probe.
Temperature probe:
Pt-1000 platinum resistor
The probe‘s accuracy in range 0 ÷ 100 °C:
for Pt1000B resistor
±0.8 °C
for Pt1000A resistor
±0.35 °C