12. pH MEASUREMENT
Before starting the measurement the meter and the pH electrode have to be
prepared for work (chapters 7 and 8 respectively). Good condition of the
electrode is crucial for obtaining the correct readings.
12.1. Measurement with automatic temperature compensation
- turn the meter on by pressing the
button;
- connect the temperature probe and the combination pH electrode to the
F
and
t
connectors respectively (Pic. 2), the symbol will be displayed;
- immerse the electrode and the temperature probe in the measured
solution. During measurements in vessels don’t touch the bottom and the
walls with the electrode. It is advisable to use an electrode stand;
- after stabilisation read the result.
Accurate laboratory measurements require using of a magnetic stirrer.
NOTE
: exceeding the measurement range is signalised with red colour of the
displayed pH value; exceeding the temperature compensation range – with
yellow colour.