VWR 12777-838 Скачать руководство пользователя страница 4

4. 

To set the high temperature alarm limit (HI), press the HI ALM SET button to advance the display until 

the desired high temperature is displayed. 

 

SETTING AMBIENT (IN) TEMPERATURE ALARM LIMITS 

1. 

Select the Ambient Temperature Display Mode by pressing the IN/OUT button located on the back of 

the unit until “IN” appears on the display. 

2. 

Press MODE button until “LO” and “HI” appears. 

3. 

To set the low temperature alarm limit (LO), press the LO ALM SET button to advance the display until 

the desired low temperature is displayed. 

4. 

To set the high temperature alarm limit (HI), press the HI ALM SET button to advance the display until 

the desired high temperature is displayed. 

ENABLE/DISABLE ALARMS 

To enable the alarm to sound when the temperature being displayed is outside the alarm limits, press the 

 

button located on the back of the unit until “

” appears on the display. Each successive press of the 

 

button will enable (

) or disable the alarm. 

ALARM SOUNDING 

With the alarm enabled (

), the unit will sound an alarm when the temperature being displayed is outside 

the alarm limits that have been set for that display mode (equal to or lower than the low alarm set point, or 

equal to or greater than the high alarm set point). If the alarm is sounding based on the low alarm limit, “LO” 

will flash on the display. If the alarm is sounding based on the high alarm limit, “HI” will flash on the display. 

The unit will continue to alarm and flash the corresponding “LO” or “HI” until the temperature being displayed 

is within the alarm limits. Once the temperature being displayed is within the alarm limits, the alarm will stop 

sounding and the corresponding “LO” or “HI” will no longer flash. 

To silence an alarm manually, press any button. When manually silenced, the corresponding “LO” or “HI” will 

continue to flash on the display until the temperature is within the alarm limits. When enabled (

), the alarm 

will only sound if the temperature being displayed is outside the alarm limits that have been set for that 

display mode (IN or OUT). Example: If the unit is displaying ambient temperature (IN) and the probe 

temperature (OUT) is outside the probe temperature alarm limits, 

no alarm will sound

RESET BUTTON 

The RESET button located on the back of the unit may be used to simultaneously reset the memory and 

alarm limits for both the ambient (IN) and probe (OUT) sensor. 

To reset the unit, use a pointed object to press the RESET button. 

When the reset button in pressed, the following will occur: 

• 

The display will show all segments for approximately 5 seconds. 

• 

The ambient (IN) and probe (OUT) minimum/maximum memory will be cleared. 

• 

The ambient (IN) and probe (OUT) alarm limits will be returned to the default value of 10°C (LO) and 

30°C (HI). 

• 

The unit will display the ambient (IN) temperature in °C. 

 

 

VWR Collection Manual ver 3, rel 5      30.10.2013 

Содержание 12777-838

Страница 1: ...CTION MANUAL North American Catalog Numbers 12777 838 89094 782 89094 768 89094 770 European Catalogue Numbers 620 1582 620 1583 620 1584 620 1585 Version 1 Issued 15th May 2014 VWR Collection Manual...

Страница 2: ...AY MODES Ambient Temperature Display Mode IN Indicated by IN appearing on the display displays the current ambient temperature and either the ambient temperature memory MIN MAX or the ambient temperat...

Страница 3: ...X RESET button to clear the maximum temperature memory CLEARING THE AMBIENT IN TEMPERATURE MEMORY 1 Select the Ambient Temperature Display Mode by pressing the IN OUT button located on the back of the...

Страница 4: ...f the alarm is sounding based on the high alarm limit HI will flash on the display The unit will continue to alarm and flash the corresponding LO or HI until the temperature being displayed is within...

Страница 5: ...lifting it out The battery cover is hinged at the bottom so that the battery compartment may be opened without completely removing the battery cover Remove the exhausted battery and replace with a AA...

Страница 6: ...nvolved in the disposal and recycling of the equipment For more information about where you can drop off your waste of equipment please contact your local dealer from whom you originally purchased thi...

Страница 7: ...i n de la temperatura ambiente IN y de la temperatura de la sonda OUT pulse el bot n IN OUT situado en la parte posterior de la unidad Para cambiar entre la visualizaci n de la memoria MIN MAX y de lo...

Страница 8: ...Pulse el bot n MODE hasta que MIN y MAX aparezca en la pantalla 3 Pulse el bot n MIN RESET para borrar la memoria de temperatura m nima 4 Pulse el bot n MAX RESET para borrar la memoria de temperatura...

Страница 9: ...unidad continuar haciendo sonar la alarma y parpadeando el mensaje LO o HI correspondiente hasta que la temperatura visualizada est dentro de los l mites de alarma Una vez que la temperatura visualiz...

Страница 10: ...sualizaci n o la aparici n de un s mbolo de pila en la pantalla indican que la pila debe ser reemplazada Abra el compartimiento de la pila presionando la leng eta en la parte superior de la tapa de la...

Страница 11: ...sonde LO HI Pour changer entre les modes d affichage de la temp rature ambiante IN et de la temp rature de la sonde OUT appuyez sur le bouton IN OUT situ l arri re de l appareil Pour changer entre l a...

Страница 12: ...de la temp rature ambiante en appuyant sur le bouton IN OUT situ l arri re de l appareil jusqu ce que l cran affiche IN 2 Appuyez sur le bouton MODE jusqu ce que l cran affiche MIN et MAX 3 Appuyez su...

Страница 13: ...nch e sur la base de la limite d alarme haute HI clignotera sur l cran L appareil continuera d clencher l alarme et faire clignoter le message LO ou HI correspondant jusqu ce que la temp rature affich...

Страница 14: ...yant sur l onglet situ en haut du couvercle du compartiment de la pile et en le soulevant Le couvercle est articul en bas pour que l on puisse ouvrir le compartiment de la pile sans enlever compl teme...

Страница 15: ...ementations locales sauf exception si le d faut du produit n est pas imputable cette non ex cution Il est recommand au client d assurer les l ments retourn s contre les risques ventuels d endommagemen...

Страница 16: ...de OUT Es zeigt die aktuelle Sondentemperatur an sowie entweder das Memory der Sondentemperatur MIN MAX oder die Alarmgrenzen f r die Sondentemperatur LO HI Um zwischen den Display Modi f r die Umgebu...

Страница 17: ...Taste MODE bis im Display MIN und MAX erscheint 3 Dr cken Sie auf die Taste MIN RESET um das Minimumtemperatur Memory zu l schen 4 Dr cken Sie auf die Taste MAX RESET um das Maximumtemperatur Memory...

Страница 18: ...t wird weiter einen Alarm abgeben und das entsprechende LO oder HI wird aufleuchten bis sich die angezeigte Temperatur wieder innerhalb der Alarmgrenzen befindet Wenn sich die angezeigte Temperatur wi...

Страница 19: ...det um diese dann herauszuziehen Die Batterieabdeckung ist unten eingeh ngt so dass das Batteriefach ge ffnet werden kann ohne die Batterieabdeckung vollst ndig zu entfernen Wechseln Sie die leere Bat...

Страница 20: ...REINBART DASS DIESE GEW HRLEISTUNG ANSTELLE JEGLICHER GEW HRLEISTUNG DER EIGNUNG UND ANSTELLE DER GEW HRLEISTUNG DER ALLGEMEINEN GEBRAUCHSTAUGLICHKEIT GILT Befolgung lokaler Gesetze und anderer Rechts...

Страница 21: ...icazione OUT viene visualizzata la temperatura effettiva della sonda e la relativa memoria della temperatura della sonda MIN MAX o i valori limite di allarme della temperatura della sonda ALTO BASSO P...

Страница 22: ...ET per azzerare la memoria dei valori minimi di temperatura 4 Premere il tasto MAX RESET per azzerare la memoria dei valori massimi di temperatura ALLARMI Entrambi i valori limite di allarme della tem...

Страница 23: ...e il comando LO o HI continuer a lampeggiare fino a quando la temperatura visualizzata sul display non rientrer nei valori limite predefiniti Se la temperatura visualizzata sul display rientra nei va...

Страница 24: ...senza dovere rimuovere completamente il coperchio della batteria Rimuovere la batteria scarica e sostituirla con una batteria alcalina AAA Assicurarsi di inserire la nuova batteria con la giusta pola...

Страница 25: ...TUISCE TUTTE LE GARANZIE DI IDONEIT E LA GARANZIA DI COMMERCIABILIT Conformit a leggi e normative locali Il cliente responsabile della richiesta e dell ottenimento delle approvazioni normative necessa...

Страница 26: ...letras OUT fora que aparecem no visor Exibe a temperatura atual da sonda e a mem ria de temperatura da sonda MIN MAX ou os limites LO HI de alarme da temperatura da sonda LO Inferior HI Superior Para...

Страница 27: ...MAX RESET para apagar a mem ria de temperatura m xima ALARMES Os limites de alarme tanto da temperaturaambiente IN como de sonda OUT podem ser definidos em incrementos de 1 grau Com o alarmearmado a...

Страница 28: ...ar e as letras correspondentes LO ou HI continuam a piscar at que a temperaturasendo exibida esteja de novo dentro dos limites de alarme Uma vez que a temperaturasendo exibida esteja dentro dos limite...

Страница 29: ...tampa Retire a bateria descarregada e coloque em seu lugar uma bateria alcalina AAA Certifique se de colocar a nova bateria de acordo com a polaridade correta conforme mostra a ilustra o no interior d...

Страница 30: ...ou outras autoriza es necess rios para executar ou usar o Produto no seu local de trabalho A VWR n o assume responsabilidades por omiss es relacionadas com este facto ou pela n o obten o da devida ap...

Страница 31: ...eland Northern Ireland VWR International Ltd VWR International Northern Ireland Ltd Orion Business Campus Northwest Business Park Ballycoolin Dublin 15 Tel 01 88 22 222 Fax 01 88 22 333 E mail sales i...

Отзывы: