4 | Explosion-Proof Refrigerator & Freezer
GP, Explosion-Proof and FMS Lab Refrigerators and Freezers
Explosion-Proof Refrigerator &
Freezer
Conventional refrigerators and freezers are not suitable for
storing flammable materials. Such units have components in
their electrical and refrigeration systems that can trigger
explosions of flammable air-vapor mixtures inside the unit
and/or in the immediate surrounding area.
VWR Scientific offers the largest explosion-proof selection in
the industry, including single and 2-door upright refrigerator
undercounter refrigerators.
Motors and thermostats are designed to prevent the arcing
which can ignite a flammable air-vapor mixtures, In addition,
each unit has wiring, splices, a thermostat, a relay and a
compressor motor safely housed within an explosion-proof
enclosure of construction suitable for use in Class I, Groups C
and D. This means that things like the surface temperature of
compressors will remain below the flashpoint of any
flammable material likely to be found in Class I, Groups C and
D, Class I Zone 1 and Group IIB locations.
All models have heavy-gauge, rigid, braced-steel construction
with a durable enamel finish. Interiors have epoxy enamel or
ABS plastic construction. Each unit is insulated throughout for
energy-efficient operation.
These VWR Scientific units will prevent potentially destructive
explosions in any facility by meeting the standards established
by Underwriters’ Laboratory, Inc., OSHA and National Fire
Protection Association for storage of hazardous materials.
Explosion-proof models are specifically designed and
engineered to prevent the triggering of an explosion inside or
outside of the unit in a Class I, Groups C and D hazardous
material environment. These models feature stainless steel
refrigeration components that are more resistant to attack
from corrosive vapors than are aluminum or epoxy-coated
aluminum.
It is important to note that Explosion-Proof models must be
connected to the power supply source by rigid metal conduit
or mineral-insulated cable with proper fittings suitable for
Class I, Groups C and D hazardous locations.In addition, the
wires inside the conduit must be sealed.
For Explosion-Proof units where flammable
materials are stored in the cooling chamber, the
cooling chamber is considered a Class I Div1 or
Class I Zone 1 hazardous location. Any
monitoring devices placed in the cooling
chamber must have an intrinsically safe rating
from an appropriate certification body, such as
UL, CSA, FM etc. (this includes battery or solar
powered devices).
Thermocouples for building monitoring systems must be
wired through an electrical barrier designed to provide
isolation against voltage and current spikes, which could
cause a spark resulting in fire or explosion. It is the your
responsibility to meet these requirements. VWR cannot assist
with the selection of devices, recommend, approve or design
any device or monitoring circuit.
MODEL 10ECEEVWA - 10.1 Cu. ft.
23.63" Wide, Two-door Refrigerator/Freezer
Features:
•
Refrigerator: 3 Adjustable Shelves
•
Freezer: 1 Adjustable Shelf
•
Hydraulic thermostat
•
Adjustable natural air-flow vent
•
Manual defrost
•
ABS plastic interior
•
White color
10ECEEVWA
Refrigerator Chamber
Dimensions
H x W x D inches (cm)
38.25" x 19.75" x 18.25"
(97.2 x 50.2 x 46.4 cm)
Freezer Chamber
Dimensions
H x W x D inches (cm)
18.25" x 18" x 11.25"
(46.3 x 45.7 x 28.6 cm)
Exterior Dimensions
H x W x D inches (cm)
59.75" x 23.63" x 28.6"
(151.77 x 60 x 76.2 cm)
Electrical Characteristics
Volts/Hz, Amps
115/60, 1.9
Refrigerator Temp. Range
°F (°C)
33.8° to 53.6°
(1° to 12°)
Freezer Temp. Range
°F (°C)
-4° to 10.4°
(-20° to -12°)
Net Weight lbs. (kg)
120 (55)
Shipping Weight lbs. (kg)
150 (68)
Note:
Amps listed are at normal run mode, starting amps
may be higher.