INTERFACE SETTINGS
All VWR-E Series Balances come complete with RS232 interface. Below you
will find the factory settings and programming instructions for the optional
settings for the Print (send), Baud Rate, Parity and Stability commands.
1.
Press
ON/OFF
Key to turn power off, press again to turn power on. When
instrument displays ---------
2.
Press PRINT
: b (Baud Rate) will appear in the display.
3.
Press
TARE
to select among 1200, 2400, 4800, 9600* Baud
4.
Press
UNIT
to select Parity settings, P NONE* will appear in the display.
5.
Press
TARE
to select among ODD, EVEN, NONE.
a)
ODD = 7 data bits with odd parity.
b)
EVEN = 7 data bits with even parity.
c)
NONE* = 8 data bit without parity.
6.
Press
UNIT
to enter PRINT/SEND settings;
n
will appear in the display.
7.
Press
TARE
to select among
a)
CON, continuous print/send (1 per second)
b)
OFF, serial data output disabled,
c)
KEY*, manual mode by pressing PRINT
d)
STB, automatic print when reading is stable
8
. To save settings
Press
ON/OFF
to turn the power off,
Press
ON/OFF
again to turn on the power and return to the weighing mode.
* Factory Default Settings