INSTALLATION &  SETUP 

- Choose a location that is free from the influences of the following: 

 

Drafts: Fans, Heat/Air Duct, Doorways or High Traffic Areas. 

 

Vibrations. 

 

Extreme  Temperature  Fluctuations:  Sunlight,  Ovens,  or  Environments 
with Wide Temperature Fluctuations 

 

Voltage  Fluctuations.    Do  Not  Share  Current  with  Other  Instruments 
that  Draw  Voltage  in  an  Inconsistent  Manner  such  as  Fluorescent 
Lights, Centrifuges and the like. 

- Place Instrument on a Level, Rigid Surface. 
- Level the Instrument. 
- Allow 30 Minutes for Warm-up. 
- Calibrate. All VWR-E Series are calibrated at the factory prior to shipment. 

Transportation of the instrument plus the differences in barometric pressure, 
humidity and ambient temperate conditions require calibration at the point of 
use.  Calibrate regularly, at least once a week, to ensure accurate weighing 
results. 

 

AC POWER CONNECTION & BATTERY OPERATION OPTION  

-  When  connecting  power  use  the  original  AC  adapter  supplied  with  the 

balance.    Using  an  unapproved  power  supply  may  damage  the  instrument 
and void the warranty. 

- Connect the plug to the instrument. 
- Ensure the power supply is compatible with your instrument AND the local 

line voltage. 

- Connect adapter to the power outlet. 
- Original Power Supply:  
 

     

VWR-1002E, 3002E, 3001E & 6001E:  VWR-10001E, 20001E 

 

  

Input:

    

100-240VAC, 50-60Hz, 0.2A   

 

120VAC, 50-60Hz,

 120mA 

  Output:

  

12VDC, 500mA 

 

 

 

10.5VDC, 500mA 

  

Polarity

:                       

Inside = 

-

, Outside = +

   

 

     

Battery:

   

6 AA   

 

   

 

 

Internal, Rechargeable

 

 

Do not mix batteries.  Replace all six batteries at the same time. If the 
instrument will not be used for an extended period of time, remove all 
batteries from the instrument.  Failure to do so may cause leakage and 
damage to the instrument. 

Page 2 

Содержание VWR-E Series

Страница 1: ...VWR E Series Balances Operation Manual...

Страница 2: ...LING 3 CALIBRATION 3 WEIGH MODE SETTINGS 4 BACKLIGHT SETTINGS 4 AUTO OFF SETTINGS 4 TARE FUNCTION 5 CARE MAINTENANCE 5 ERROR MESSAGES TROUUBLESHOOTING GUIDE 6 INTERFACE SETTINGS 7 RS232 DATA INTERFACE...

Страница 3: ...ument Avoid exceeding overloading the weigh capacity or dropping samples onto the weigh pan Weight overload and shock may damage the instrument and void the warranty When utilizing the battery operati...

Страница 4: ...ation at the point of use Calibrate regularly at least once a week to ensure accurate weighing results AC POWER CONNECTION BATTERY OPERATION OPTION When connecting power use the original AC adapter su...

Страница 5: ...ss Tare to zero balance if required 3 Remove all items from weigh pan 4 Press CAL Key The appropriate calibration weight for your balance see chart below will flash on the display 5 Place calibration...

Страница 6: ...ON will appear on the display To toggle between Backlight On L_ON and Backlight Off L_OFF Press TARE KEY Press ON OFF to turn the power off Press ON OFF again to turn on the power and the scale retur...

Страница 7: ...ng is 0 before placing a sample on the weigh pan If not press TARE KEY to display 0 CARE MAINTENANCE Unplug the instrument from the power source and remove all batteries from the battery compartment W...

Страница 8: ...at Pan Sub Pan and Calibration Calibrate ERR 5 Internal Signal Weigh Pan Alignment Unstable Ensure Weigh Pan Draft Ring Placement are Correct Nothing Touching Unstable Power Check Power Supply Transfo...

Страница 9: ...4 Press UNIT to select Parity settings P NONE will appear in the display 5 Press TARE to select among ODD EVEN NONE a ODD 7 data bits with odd parity b EVEN 7 data bits with even parity c NONE 8 data...

Страница 10: ...Polarity 2BH Positive 2DH Negative DATA 12 345 20H 20H 20H 20H 31H 32H 2EH 33H 34H 35H UNIT g 67H 20H 20H 20H 20H oz 6FH 7AH 20H 20H 20H lb 6CH 62H 20H 20H 20H dwt 64H 77H 74H 20H 20H ozt 6FH 7AH 74H...

Страница 11: ...trument that has been exposed to biological or hazardous material contamination for replacement credit repair or disposal DISPOSAL Please consider the environment when disposing of your instrument and...

Страница 12: ...esign within Certain Voltage Limits Applicable European Standards EN 60950 Safety Requirements for electrical equipment for measurement control and laboratory use Part1 General requirements Note Modif...

Страница 13: ...7 x 7 1in VWR 10001E 20001E 200 x 215mm 7 9 x 8 5in Stabilization Time 3 Seconds Operating Temperature 5C to 40C 41 104F Power Requirements Input 100 240VAC 50 60 Hz 0 2A Output 12VDC 500 mA Polarity...

Страница 14: ...t Off Mode Automatic Calibration Low Battery Indicator Sixteen Weigh Modes Overload Protection Stainless Steel Weigh Pans RS232 Communication Two Year Warranty INSTRUMENT DIAGRAM BACK SIDE AC ADPTER I...

Страница 15: ......

Страница 16: ...dicated by the date given below VWR International LLC reserves the right to make changes to any or all of the specifications features or design of the instruments at any time and without notice Date A...

Отзывы: