ONFIGURATION
This entry page is used to change the time configuration of the EX1200A. As the EX1200A has no
notion of “wall clock” or calendar time a
nd no battery-backed clock or any other mechanism to
retain time between reboots and power-cycles, the time must be set by the user. By default, the
instrument’s time and date at power
-up are midnight, January 1, 1999 (as defined by the
microcontroller). Three time source options are provided:
PTP
(Precision Time Protocol or
IEEE 1588 synchronization),
SNTP
(simple network time protocol), or
manual
. The default time
source is PTP.
Time Zone
:
allows the user to select an appropriate time zone.
Time Source
: allows the user to set the EX1200As date by entering the month (MM), day
(DD), and year (YYYY), respectively.
Set Time
: for the
PTP
and
Manual Time Source
, allows the user to set the EX1200As time
by entering the hour (HH), minute (MM), and second (SS), respectively. If the EX1200A is
set as an IEEE 1588 Master, the hour should be entered using the 24-hour format.
NTP Servers
: for the
SNTP Time Source
, allows the use to add/remove/save NTP servers to
which the EX1200A will synchronize its time.
Time Status
: this section of the web page provides the mainframe’s current time
configuration settings.
Using PTP Time
To configure the EX1200A to use PTP time as the clock source:
1)
Set
Time Source
to
PTP
. The page will change so that it can be configured.
2)
Click
Submit
to start using PTP time.
Содержание Ametek EX1200A Series
Страница 17: ...VTI Instruments Corp 16 EX1200A Series Introduction EX1200A SERIES 1 U FULL RACK MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 18: ...www vtiinstruments com EX1200A Series Introduction 17 EX1200A SERIES 3 U MAINFRAME DIMENSIONAL DIAGRAM ...
Страница 33: ......
Страница 51: ...VTI Instruments Corp 50 EX1200A Series Preparation for Use ...
Страница 64: ...www vtiinstruments com EX1200A Series Preparation for Use 63 ...
Страница 65: ......
Страница 73: ......
Страница 107: ...VTI Instruments Corp 106 EX1200A Triggering ...
Страница 109: ...VTI Instruments Corp 108 EX1200A Triggering FIGURE 6 1 EX1200A SERIES TRIGGER MODEL ...
Страница 143: ...VTI Instruments Corp 142 EX1200A Web Page Operation ...
Страница 145: ...VTI Instruments Corp 144 EX1200A Series Programming ...
Страница 155: ......
Страница 161: ......