  –  Allan  variance  of  "

fx"  and  "

fy"  input  signals  and  the 

measuring channel CHANNEL 1. 

 

 

 

 

 

 

 

 

 

Figure 5 

 

Instability  of  the  measuring  channel  CHANNEL 1  – 

1

1

1

,

,

c

c

N

c

y

("noise  floor") 

can be calculated if the same signal is connected to both "

fx" and "

fy" inputs.    

The "

fy" input signal instability estimate is biased and includes the instability 

(

systematic  error

)  of  the  reference  signal  and  the  measurement  channel.  When  using 

this  method  you  need  to  be  sure  that  instability  of  the  reference  signal  and 

measurement channel is much less than that of measured signal.  

Signal 

Source 

Y

1

 

 

RS-232 interface 

cable 

(included 

with 

the 

instrument 

set 

composition) 

UTP cable 

8P8C plug 

Personal Computer 

 

 

Multichannel frequency comparator VCH-315M 

 

fy

                                   

 

 

fx 

 

CHANNEL 1 

 

RS-232 

LAN 

~220 V 50 Hz  40 V·A 

 22...30 V, 35 W 

POWER SUPPLY

 

PROCESSOR

 

USB 

Power connecting cord  

AC power connecting cord 
(connector  2PMT14КПН4Г1В1) 

USB interface cable 

 

fy

                        

 

 

fx 

 

CHANNEL 2 

 

CHANNEL 3 

 

CHANNEL 4 

 

CHANNEL 8 

 

 

fy

                        

 

 

fx 

 

 

fy

                        

 

 

fx 

 

 

fy

                    

 

 

fx 

 

Connecting  

cables,  50 Ohm 

 

Connecting  

cables,  50 Ohm 

 

Signal 

Source 

X

1

 

Signal 

Source 

Y

2

 

Signal 

Source 

X

2

 

Signal 

Source 

Y

3

 

Signal 

Source 

X

3

 

Signal 

Source 

Y

4

 

Signal 

Source 

X

4

 

Signal 

Source 

X

8

 

Signal 

Source 

Y

8

 

Содержание VCH-315M

Страница 1: ...k Company MULTICHANNEL FREQUENCY COMPARATOR VCH 315M Operation Manual 411146 039OM Osharskaya str 67 Nizhny Novgorod 603105 Russia tel fax 7 831 421 02 94 http www vremya ch com Инв подл Подп и дата Взам инв Инв дубл Подп и дата ...

Страница 2: ...2 Key applications 5 3 3 Product composition 7 3 4 Product specifications 8 3 5 Basic configuration and operation 11 4 Operating procedure 15 4 1 Safety requirements 15 4 2 Location of control and connection systems 15 4 3 Preparing the device for use 17 4 4 Operating procedure 21 4 5 Measurement modes 22 5 Troubleshooting 34 ...

Страница 3: ...3 1 List of abbreviations ADEV Allan deviation PC personal computer RS232 serial port Multichannel frequency Comparator VCH 315M device ...

Страница 4: ... three wire power cable two poles and ground included in the device kit 2 5 To avoid the influence of static electricity all further connections must be made only with the grounded device 2 6 You must avoid contact with current carrying parts during repair as it may result in shock 2 7 Before touching any components or wiring always unplug the device 2 8 All repair and maintenance of the device mu...

Страница 5: ...rameters and control commands are transferred via RS 232 interface USB or local area network LAN The external view of the device is given in Fig 1 3 2 Key applications The main applications of the device crystal and quantum frequency standards production and checking scientific research in metrology as an integral part of the storage systems as an integral part of automated measurement systems Bui...

Страница 6: ...6 Figure 1 External view of the device ...

Страница 7: ...er connecting cord 1 4 Interface RS 232 cable 685622 003 1 5 Interface cable USB 2 0 AM BM 1 8M 1 6 Power splitter ZFSC 2 1W S 2 7 RF interconnecting cable 685670 372 4 8 RF interconnecting cable 685670 376 2 9 Multichannel frequency comparator VCH 315M Installation disc 1 10 Multichannel frequency comparator VCH 315M Operating Manual 1 11 Multichannel frequency comparator VCH 315M User guide 1 ...

Страница 8: ... or similar Operating system Microsoft WindowsTM XP Vista 7 8 10 RAM 512 MB 3 4 2 Input signal parameters nominal frequencies 5 10 100 MHz RMS voltage 1 0 0 2 V at 50 1 Ohm relative frequency difference 5 0 10 9 to 5 0 10 9 3 4 3 Number of measuring channels 8 eight modules with two inputs for the measured signals fх fy 3 4 4 Averaging time range from 1 s up to 1 day 3 4 5 Noise passband 3 Hz 3 4 ...

Страница 9: ...els mode cross ADEV 1 s 6 0 10 14 2 0 10 14 10 s 2 0 10 14 3 0 10 15 100 s 3 0 10 15 1 0 10 15 1 h and more 3 0 10 16 2 0 10 16 Note frequency instability introduced by the comparator is guaranteed after warm up time according to s 3 4 6 3 4 8 The device provides continuous non stop operation in working conditions with all specifications preserved 3 4 9 Power supply voltage AC power supply 220 22 ...

Страница 10: ... AC power supply not more 40 VA DC power supply not more 35 W 3 4 12 Ambient temperature operating 5 to 40 С non operating 50 to 50 С 3 4 13 Weight no more than 12 kg net 16 kg in factory package 3 4 14 Dimensions 133 483 325 mm ...

Страница 11: ...e measurements to the computer where all the data is processed by a special program that gives the characteristics of the input signals instability in digital and graphical forms 3 5 2 Description of the circuit diagram Circuit diagram illustrating the operating principle of the device is shown in Figure 2 The device consists of these main parts eight identical frequency comparators channels proce...

Страница 12: ...on of the frequency comparators by commands from the special program and also controls the indicators Transmission of measurement data to the computer and reception of control commands is carried out by the processor via RS 232 USB or LAN interfaces Tested signals 5 10 100 MHz SIGNALS AND POWER BUS INDICATION SIGNAL DRIVER fy fx FREQUENCY COMPARATOR 1 CHANNEL 1 tyx fy fx FREQUENCY COMPARATOR 8 CHA...

Страница 13: ...Ds turning on and off at a frequency of about 1 Hz The front panel also contains the AC 220 V power indicator and the external DC source indicator Indication signal driver generates indication signals and sends them to the front panel indicators by the processor command Power supply and internal information exchange between the device units are carried out via the signals and power bus The device ...

Страница 14: ... automatic update of charts and statistical characteristics a wide range of computed parameters mean relative frequency difference and frequency drift by least squares method standard deviation Allan deviation Hadamard deviation maximum time interval error time interval deviation power spectral density of phase and frequency noise calculation of the statistical characteristics with overlap method ...

Страница 15: ...is prohibited The device is connected to the AC via three wire power cable two poles and ground included in the device composition ATTENTION Usage of any other power plugs is STRICTLY PROHIBITED Operation of the device must comply with the electrical safety regulations in force at the operating facility 4 2 Location of control and connection systems Connectors indicators and control elements on th...

Страница 16: ...16 Figure 3 Connectors indicators and control elements on the front and rear panels ...

Страница 17: ... Normal operation indicator of the power supply unit 13 Ground terminal 4 3 Preparing the device for use 4 3 1 Operational limitations It is recommended to place the device in a thermostabilized room with limited access to personnel The place for the device should be chosen taking into account the dimensions of the device 133 483 325 mm and the need for free convection of air through the ventilati...

Страница 18: ... before starting Note the location of the controls and connections Figure 3 for their purpose Table 3 Ensure reliable grounding of the device The protective grounding terminal should be connected to the grounding conductor before other connections The fasteners of the ground terminal and the conductors must be securely fixed While the power is off connect the device to the computer in one of the f...

Страница 19: ...urce Make sure that the POWER NORM and 22 30 V 35 W indicators on the power module panel are lit The 22 30 V indicator on the power unit panel lights up only if there is a DC voltage on the 22 30 V 35 W connector To connect an external power source use a power cable with a 2PMT14КПН4Г1В1 2PM14 4TKQB2 type connector Pin 1 of the connector must be connected to the positive pole of the DC power suppl...

Страница 20: ...nt data of frequency comparators counters Analyser User Guide Before using the USB interface of the device install a PC driver that creates a virtual serial port when connecting the comparator via the USB interface following the instructions in the Multichannel frequency comparator VCH 315M User guide To control the device via the Ethernet install the special software following the instructions in...

Страница 21: ...use with the guaranteed frequency instability given in Table 2 after a warm up period Start the measurements in the Multichannel frequency comparator program following the instructions in the Multichannel frequency comparator VCH 315M User guide You can view the instability characteristics of the frequency of measured sinusoidal signals during or after the end of the measurements following the ins...

Страница 22: ...umption that frequency fluctuations of all signals and measuring channel are independent Table 3 Available functions for two oscillators one channel mode Number Calculated function Mathematical expectation 1 Relative frequency difference measured for CHANNEL 1 N c N x N y y y y 1 1 1 2 Standard deviation measured for CHANNEL 1 2 1 2 1 2 1 c x y 3 Allan deviation measured for CHANNEL 1 2 1 2 1 2 1 ...

Страница 23: ... signal and measurement channel is much less than that of measured signal Signal Source Y1 RS 232 interface cable included with the instrument set composition UTP cable 8P8C plug Personal Computer Multichannel frequency comparator VCH 315M fy fx CHANNEL 1 RS 232 LAN 220 V 50 Hz 40 V A 22 30 V 35 W POWER SUPPLY PROCESSOR USB Power connecting cord AC power connecting cord connector 2PMT14КПН4Г1В1 US...

Страница 24: ... the measurement results And the measurement error introduced by the channel can be reduced using cross correlation processing The channels noise can be estimated along with the measured signals instability Accuracy of estimations depends on the measurement time interval Table 4 shows mathematical expectation of calculated functions for the measuring channels pair CHANNEL 1 and CHANNEL 2 These fun...

Страница 25: ... 1 2 1 2 1 c x y 5 Standard deviation measured in CHANNEL 2 2 2 2 1 2 1 c x у 6 Sum standard deviation CHANNEL 1 and CHANNEL 2 2 2 2 1 c c 7 Standard cross deviation CHANNEL 1 and CHANNEL 2 2 1 2 1 x y 8 Allan deviation measured in CHANNEL 1 2 1 2 1 2 1 c x y 9 Allan deviation measured in CHANNEL 2 2 2 2 1 2 1 c x y 10 Sum Allan deviation CHANNEL 1 and CHANNEL 2 2 2 2 1 c c 11 Allan cross deviatio...

Страница 26: ...t composition UTP cable 8P8C plug Personal Computer Multichannel frequency comparator VCH 315M fy fx CHANNEL 1 RS 232 LAN 220 V 50 Hz 40 V A 22 30 V 35 W POWER SUPPLY PROCESSOR USB Power connecting cord AC power connecting cord connector 2PMT14КПН4 USB interface cable fy fx CHANNEL 2 CHANNEL 3 CHANNEL 4 CHANNEL 8 fy fx fy fx fy fx Signal source Y1 Signal source X1 Connecting cables 50 Ohm Connecti...

Страница 27: ... of the instability of each individual oscillator reduced systematic error caused by instability of reference signal and measuring channel RS 232 interface cable included with the instrument set composition UTP cable 8P8C plug Personal Computer Multichannel frequency comparator VCH 315M fy fx CHANNEL 1 RS 232 LAN 220 V 50 Hz 40 V A 22 30 V 35 W POWER SUPPLY PROCESSOR USB Power connecting cord AC p...

Страница 28: ...CHANNEL 2 N c N c N y N y y y y y 3 1 3 1 5 Sum relative frequency difference CHANNEL 2 and CHANNEL 3 N c N c N y N y y y y y 1 2 1 2 6 Sum relative frequency difference CHANNEL 3 and CHANNEL 1 N c N c N y N y y y y y 2 3 2 3 7 Standard deviation measured in CHANNEL 1 2 1 2 3 2 1 c y y 8 Standard deviation measured in CHANNEL 2 2 2 2 1 2 2 c y y 9 Standard deviation measured in CHANNEL 3 2 2 2 2 2...

Страница 29: ...rd cross deviation CHANNEL 3 2 3 2 2 c y 21 Standard cross deviation CHANNEL 1 2 1 2 3 c y 22 Allan deviation measured in CHANNEL 1 2 1 2 3 2 1 c y y 23 Allan deviation measured in CHANNEL 2 2 1 2 1 2 2 c y y 24 Allan deviation measured in CHANNEL 3 2 1 2 2 2 3 c y y 25 Sum Allan deviation CHANNEL 1 and CHANNEL 2 2 2 2 1 2 2 2 3 c c y y 26 Sum Allan deviation CHANNEL 2 and CHANNEL 3 2 3 2 2 2 3 2 ...

Страница 30: ...7 give systematic error caused by instability of reference signal in both channels and instability of measuring channels Formulas 14 15 17 18 20 21 29 30 32 33 35 36 give systematic error caused by instability of measuring channels Formulas 13 16 19 28 31 34 give non biased estimations We can measure without systematic error only Y1 signal in measuring channels CHANNEL 1 CHANNEL 2 only Y2 signal i...

Страница 31: ...connect the device and sine signal source with frequency distribution amplifier in the example for CHANNEL 1 2 or with power splitters in the example for CHANNEL 4 8 as shown in Figure 8 After 6 hour warm up time start measuring frequency instability Allan deviation introduced by measuring channels for averaging times 1 s 10 s 100 s 1000 s 1 h for 16 18 hours Note Use RF interconnecting SMA BNC ca...

Страница 32: ...ed in CHANNEL 2 2 2 c 6 Sum standard deviation CHANNEL 1 and CHANNEL 2 2 2 2 1 c c 7 Standard cross deviation CHANNEL 1 and CHANNEL 2 0 8 Allan deviation measured in CHANNEL 1 2 1 c 9 Allan deviation measured in CHANNEL 2 2 2 c 10 Sum Allan deviation CHANNEL 1 and CHANNEL 2 2 2 2 1 c c 11 Allan cross deviation CHANNEL 1 and CHANNEL 2 0 Notes Formulas 4 5 6 8 9 10 give systematic error caused by ch...

Страница 33: ...32 LAN 220 V 50 Hz 40 V A 22 30 V 35 W POWER SUPPLY PROCESSOR USB Power cord AC power cord connector 2PMT14КПН4 USB interface cable fy fx CHANNEL 2 CHANNEL 3 CHANNEL 4 CHANNEL 8 fy fx fy fx fy fx OUT A1 OUT A2 OUT A3 OUT A4 Frequency Distribution Amplifier IN A Signal Source X2 1 2 Power splitter ZFSC 2 1W s Signal Source X1 1 2 Power splitter ZFSC 2 1W s Connecting cables 50 Ohm ...

Страница 34: ...er cord fault Repair or replace the cable Device malfunction Factory repair Front panel CHANNELS 1 2 3 4 5 6 7 8 indicators not flashing at 1 Hz when input signals are present No input signal Connect input signal Connecting cable fault Repair or replace the interconnecting cable Device malfunction Factory repair RS 232 communication failure RS 232 interface cable fault Repair or replace the RS 232...

Страница 35: ...35 Problem Probable cause Solution Notes LAN communication failure Ethernet error Repair or replace the Ethernet cables Test the PC Ethernet interface Device malfunction Factory repair ...

Отзывы: