Interferential Stimulation:

Pulse amplitude:  

0 - 25V peak 

Pulse shape:     

 

With four (4) electrodes – symmetric, biphasic 
square wave 
With two (2) electrodes – symmetric, biphasic 
rectangular wave

Pulse width:   

With four (4) electrodes – 125 μs per phase

 

 With two (2) electrodes – 5 μs to 125 μs per 
phase 

Output Confi guration: 

Quadrapolar:   

 

 

Ch. 1 delivers a fi xed-frequency carrier pulse 
train. Ch. 2 delivers a user-adjustable fre-
quency that consists of the carrier frequency, 
plus a beat frequency. Four (4) electrodes are 
required. 

Bipolar:   

 

 

Ch. 1 delivers a pulse train that is premodu-
lated inside the unit. Ch. 2 is not used and 
has no stimulation pulses. Two (2) electrodes 
are required. 

Carrier frequency:  

 

4000 Hz 

Adjustable frequency: 
      Continuous mode range:  

4001 Hz - 4150 Hz, in 10 Hz increments
(beat frequency of 1 - 150 Hz)

 

 

Sweep modes 1 

1 - 10 Hz over a 30-sec period

        

 

1 - 150 Hz over a 30-sec period

     

 

80 - 150 Hz over a 30-sec period

     

 

1 - 10 Hz over a 15-sec period

        

 

1 - 150 Hz over a 15-sec period

     

 

80 - 150 Hz over a 15-sec period

Frequency shift mode range: 

4000.6 Hz - 4240.0 Hz 

Содержание Orthostim 4

Страница 1: ...Patient Manual OrthoStim4 Rehabilitative Electrotherapy from VQ OrthoCare...

Страница 2: ...equired regarding this product and its operating instructions This patient manual is proprietary and con dential No part of this document may be reproduced or transmitted in any form or by any means e...

Страница 3: ...es 6 Accessories 8 Starting Your Treatment Sessions 10 Normal Running Screens 13 After Completing Your Treatment Sessions 17 Electrode Use and Care 18 Cradle Modem 19 Care and Maintenance 22 Technical...

Страница 4: ...of this technology to make modi cations to your settings and to assist all of your healthcare partners in providing the most personalized service possible In order to gain maximum bene t from your Ort...

Страница 5: ...re 2 Power Packs The OrthoStim4 system also includes 1 carrying case with removable shoulder pouch 1 Patient Manual this booklet 2 additional packages of non sterile reusable electrodes depending on y...

Страница 6: ...ferential Stimulation can be used in the following applications Relieves acute pain Relieves chronic pain Relaxation of muscle spasms Maintaining and increasing range of motion Increases local blood c...

Страница 7: ...ealing process Over the menstruating or pregnant uterus Over areas of the skin which lack normal sensation Some patients may experience skin irritation or hypersensitivity due to electrical stimulatio...

Страница 8: ...sor down in a selection list Rotate stimulation to the left in vector mode The button has three functions It can be used to Increase the intensity of stimulation through pressing and releasing the but...

Страница 9: ...l displays as white and starts to ash and the device will beep once a minute Replace the alkaline Power Packs as soon as possible When battery power is suf ciently depleted the device will shut off au...

Страница 10: ...pter or by placing the OrthoStim4 in the cradle Power Packs The unit should be turned off when switching from one power source to the other Additional Power Packs can be ordered from VQ OrthoCare s Pa...

Страница 11: ...Press PAUSE to resume treatment Cradle Modem The OrthoStim4 system includes a cradle that contains a modem that enables the download of updates and information to from the OrthoStim4 through a landlin...

Страница 12: ...not to cut or break the skin DO NOT SHAVE THE SKIN IMMEDIATELY PRIOR TO STIMULATION because shaving will create small cuts in the surface layer of skin CAUTION DO NOT apply lotions oils or other oint...

Страница 13: ...N OFF button the OrthoStim4 will start its self test When the self test is completed you will see Select Protocol at the top of the screen and a number 1 ashing in the middle of the screen along with...

Страница 14: ...quipment 7 Select the preset as directed by your physician or therapist by pressing the or button until the number indicated for the protocol is shown in the middle of the screen Press S to select 8 I...

Страница 15: ...The center top of the screen displays the protocol number you are using The upper right corner displays the type of protocol currently running e g IF NM or HV The middle of the screen displays the amp...

Страница 16: ...ectoring During an interferential IF treatment the OrthoStim4 allows you to vector the stimulation Vectoring allows you to move the stimulation pattern so that you feel more stimulation under the pair...

Страница 17: ...s to adjust NOTE Your OrthoStim4 may be programmed slightly different than shown in the screen diagrams above In NM neuromuscular mode the screen may read Red and Black instead of Gray and Black In th...

Страница 18: ...atment the following screen will be displayed 1 HV 40V PAUSE Pause 00 39 59 Normal Running Screens continued HIGH VOLT PULSED CURRENT treatment screen If you are running a high volt treatment the foll...

Страница 19: ...ner in the resealable pouch X SL SLEEP 00 29 59 Some protocols may include a SLEEP segment During this segment the device does not provide stimulation but the timer is counting down to the next treatm...

Страница 20: ...for treatment Always make sure electrodes are placed at least one inch apart from each other To remove the electrodes Grasp the electrode by the edge and gently remove it from the skin Return it to th...

Страница 21: ...place electrodes that do not stick uniformly or cause discomfort The OrthoStim4 Cradle is equipped with a modem that allows VQ OrthoCare to make modi cations to the device from a remote location and t...

Страница 22: ...acks green The battery is fully charged The Patient Technician who instructed you in the use of the OrthoStim4 will set up the modem and connect it to your telephone line If you need to reinstall the...

Страница 23: ...dle weekly to provide your physician with your current treatment data If the device does not connect in 90 days the prescription time will expire and you will see the following screen To manually down...

Страница 24: ...terference with normal activity These wires only need to be cleaned if they become dirty Wipe with a damp cloth and dry thoroughly Storage When not in use store the OrthoStim4 in its carrying case to...

Страница 25: ...deral Communications Commission FCC Part 68 rules and requirements adopted by the Administrative Council for Terminal Attach ments ACTA The cradle modem is designed to be connected to a compatible mod...

Страница 26: ...it yourself If the equipment is causing harm to the telephone network the telephone company may request that you discontinue use of the cradle modem until the situation is resolved This equipment mus...

Страница 27: ...25...

Страница 28: ...r each modality 30mA for IF 60 mA for NM 200mA for HV 60mA for PDC to reduce current density Garment can be used with conductive garments This setting will disable the load sensor Electrical tolerance...

Страница 29: ...s of the carrier frequency plus a beat frequency Four 4 electrodes are required Bipolar Ch 1 delivers a pulse train that is premodu lated inside the unit Ch 2 is not used and has no stimulation pulses...

Страница 30: ...beat frequency gradually changes from 60 of user setting to 160 of user setting over a six second period Over the next six seconds the beat frequency gradually changes back to 60 of user setting The...

Страница 31: ...quency modulation S1 1 20 Hz over a 45 sec time frame S2 1 120 Hz over a 45 sec time frame S3 80 120 Hz over a 45 sec time frame Coordination modes Continuous Stimulation is provided to all active out...

Страница 32: ...t waveform in the HVPC stimulation mode at maximum amplitude 330V The waveform is shown with a purely resistive load of 500 ohms The signal repeats at a period of time equal to the recip rocal of the...

Страница 33: ...utput waveform in the single channel premodu lated IF stimulation mode at maximum amplitude 50mA The waveform is shown with a purely resistive load of 500 ohms The two signals created in two channel m...

Страница 34: ...the output waveform in the two channel IF stimula tion mode at maximum amplitude 50mA The waveform is shown with a purely resistive load of 500 ohms Ch 1 is shown Ch 2 is identical except that the pe...

Страница 35: ...timulation mode at maximum amplitude 100mA The waveform is shown with a purely resistive load of 500 ohms The stimulation on each channel is the same The signal on Ch 1 occurs 180 degrees out of phase...

Страница 36: ...la tion mode at maximum amplitude 100mA The waveform is shown with a purely resistive load of 500 ohms The stimulation on each channel is the same The signal on Ch 1 occurs 180 degrees out of phase wi...

Страница 37: ...imulation This drawing shows the 277 Hz output waveform in the pulsed direct current stimulation mode at maximum amplitude 100mA The waveform is shown with a purely resistive load of 500 ohms The stim...

Страница 38: ...imulation This drawing shows the 222 Hz output waveform in the pulsed direct current stimulation mode at maximum amplitude 100mA The waveform is shown with a purely resistive load of 500 ohms The stim...

Страница 39: ...ls to correct the problem contact Patient Care Device beeps and the screen says Error Check your lead wires and electrodes In most cases this screen will appear when something becomes disconnected e g...

Страница 40: ...he set up mode where you set the intensity for each treatment and the actual treatment program has not yet started there will not be a timer at the bottom of the screen even though you will feel the s...

Страница 41: ...39 Notes...

Страница 42: ...40 Notes...

Страница 43: ...OrthoStim4 unit found to have become defective within the warranty period This warranty does not apply to accessories speci cally lead wires electrodes batteries tapes gel carrying case and AC Adapter...

Страница 44: ...18011 Mitchell South Irvine CA 92614 Call 800 266 6969 Fax 800 821 8012 www vqorthocare com customerservice vqorthocare com 2010 VisionQuest Industries Inc VQO021578REVA...

Отзывы: