Troubleshooting (cont.)
Problem: Battery will not hold a charge.
Your BioniCare Generator may display a “Low Battery” indication. Keep track of
how long the battery lasts during one treatment session. If the battery lasts eight
hours or more, the battery is functioning normally. Remember to charge both
batteries initially for at least 4 hours; be sure the battery charger is not plugged
into an outlet controlled by a light switch. In order to hold their maximum charge,
the batteries must be cycled (fully charged and fully drained) 4 to 5 times. If you’re
travelling overseas be sure to get a converter adapter for the charger.
Problem: Display is blank and the BioniCare Generator shuts off.
If your Generator shuts off, you may have accidentally turned it off or unplugged
the Wire. If you can turn on the Generator normally then this is most likely the
problem. If the Generator does not turn back on, it might mean the battery has
completely discharged and has turned off normally. Plug in a charged battery
and check the battery percentage. A fully charged battery should say “100%”.
One last thing; be sure to hold down the On/Off button for about one second; if
the button is not pressed long enough the Generator will not turn on. The best
way to know you’re doing it right is to wait for BioniCare logo to display, then
let go of the On/Off button.
BioniCare Generator
The case of the BioniCare Generator may be cleaned with a cloth and cotton
swab. The BioniCare Generator should be turned “OFF” and not connected to the
Wire when cleaning. Do not use liquids for cleaning; the BioniCare Generator is
not waterproof. The BioniCare Generator case must not be opened by removing
the screws, because there are no operator serviceable components inside.
Always store the BioniCare Generator in a clean, dry area between 0° and 140°F.
•
Do not use the BioniCare Generator with a battery other than the
original battery supplied or an exact replacement battery.
•
Do not use any Wire or cord if the insulation is damaged or if the
conductive strands are exposed.
•
Do not immerse the BioniCare Generator, batteries, battery charger
or Wires in water or any other liquid.
•
Do not wear and use the Device while bathing, swimming or any
other activity that would get the Device wet.
•
Do not use the BioniCare battery charger with extension cords or
adaptor plugs.