![VPD 2in1 MICRO Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/vpd/2in1-micro/2in1-micro_user-manual_1052368055.webp)
55
Clinical accuracy
The accuracy of the 2in1. MICRO
Blood Glucose Monitoring System was
assessed by comparing blood glucose
results obtained by patients with those
obtained using a YSI 2300 Glucose
Analyzer.
System accuracy result for glucose concentrations
< 100 mg/dL (5.5 mmol/L)
System accuracy result for glucose concentrations
≥
100 mg/dL (5.5 mmol/L)
within ±5 mg/dL (0.28 mmol/L)
94.0%
within ±10 mg/dL (0.56 mmol/L)
100.0%
within ±15 mg/dL (0.83 mmol/L)
100.0%
within ±5%
88.0%
within ±10%
100.0%
within ±15%
100.0%
within ±20%
100.0%