Control Solution Testing
46
STEP 8
Compare the result displayed on the
screen to the control solution range
printed on the test strip vial. A single
control solution test that is with range
is sufficient to show that your meter
and test strips are working as intended.
If the results you get are not within this
range, the meter and strips may not be
working properly. Repeat the control
solution test. If you get a second control
solution result, call the local distributor.
Do not use the meter if it is not testing
within range.
STEP 10
Remove the used test strip and touch
”save” button to store test result.
Control Test