4. INSTALLATION AND INITIAL START-UP
UNPACKING
CONNECTIONS
SETTING THE TIME TO AUDIBLE SIGNALS
IMPORTANT!
Carefully remove modules from packages;
Make sure that the mains voltage is ~220V and
the socket is reliably grounded.
Connect pump supply cable to the terminal with
label "VACUUM PUMP" on the rear panel;
Connect the power supply cable to the terminal
with label "POWER" on the rear panel;
Connect vacuum hose to the vacuum fitting with
label "VACUUM” on the rear panel.
The net switch is on the rear panel, on the right.
Put the furnace in a position which enables its
easy engaging turning off from the master
switch as well as from the contact.
You can set the duration of the sound signaling at
the end of the program. For this purpose it is
necessary to perform the following:
Turn off the furnace;
Press RUN and STOP keys simultaneously;
Hold the buttons until the next message appears:
END CYCLE SOUND
00015
The current value can be changed by turning the
multifunction knob
MFK
. The new value is stored
by pressing the button
PROG
.
After turning the oven on display for a few
seconds will show SELF TESTING. Do not release
the buttons until this sign stands on display.