Ceramic Master E10 en v2
16
Transfer of a LOG file from the Ceramic Master E10
furnace to PC.
Switch off the furnace; keep the buttons RUN and STOP
depressed simultaneously and switch on the power
supply.
Wait to appear message:
TERMO CORRECT: - - - -
Press the STOP button.
On the display following message will appear:
STOP SOUND: YES (N0)
Press ESC button.
Following question will appear:
Programs <- PC?
Press the ESC button and this question will appear on
the furnace display:
Programs -> PC?
Press the ESC button and following question will appear
on the furnace display:
Log File to PC?
From the PC start the program Teraterm Pro.
Enter the SETUP/Serial Port menu and perform
following settings:
Com x
-
Serial port number
Baud Rate
-
9600
Data
-
8 bit
Parity
-
none
Stop
-
1 bit
Flow Control
-
none
Transmit Delay -
0ns
1000ms
Go to the menu
File/Log
and perform following
settings:
Options
-
binary
Enter the file name where the coded data from the
furnace to be saved and select Open.
Press the PROG button of the furnace.
After the end of the process you can continue to work
with the furnace.
Close the log file in the PC by going to the
Teraterm Log
window and pressing the CLOSE button.