R
EMOTE
O
PERATION
Power Analyzer User Manual
© Voltech 2010
Page 43
":INT:CLK:DATE xxxxxxxx"
Sets the start date for the integrator when
configured for Clock Start Method. Start
date sent in current date format;
xxxxxxxx means dd-mm-yyyy or mm-dd-
yyyy or yyyy-mm-dd according to the Date
Format settings in the Main Menu ->
System Configuration -> Clock -> Date
Format.
":INT:CLK:DUR <value>"
Sets the duration of the integrator, in
minutes, when configured for Clock Start
Method. (1.0
≤ <value> ≤ 1,000,000)
":GRA:INT:WFM <value>"
Configure integrator graph to display a
waveform.
<value> = 0
→ Whrs
<value> = 1
→ Ahrs
<value> = 2 → Vahrs
<value> = 3 → VArhrs
<value> = 4 → Watts
<value> = 5 → VA
<value> = 6 → Var
<value> = 7 → Amps
<value> = 8 → Volts
":GRA:INT:SHW"
Change display to show integrator graph.
Return to results screen with the
:DSP:Z04 command.
":GRA:INT:SCL <value>"
Configure vertical scale for the selected
result in the integrator graph. (0.0999
≤
<value>
≤ 100,000)
":GRA:INT:DUR <value>"
Configure the horizontal scale (duration),
in minutes, when configured for Manual
Start Method. (1.0
≤ <value> ≤ 1,000,000)
":MOD:NOR”
Sets normal mode.
":MOD:INR”
Sets inrush mode.
“:MOD:INT”
Sets integrator mode.
":MOD:SBY”
Sets standby power mode.