6.Technical Specifications
36
[1]
Specifications are for 30-minute warm-up, "Low" measurement rate and calibration temperature
18
℃
- 28
℃
.
[2]
10% over range on all ranges, except 1,000 V DCV, 750 ACV, 10 A DCI, 10 A ACI, 50 MΩ resistance, and
50 mF capacitance.
[3]
For each additional volt over ± 500 VDC add 0.02 mV of error.
[4]
Specifications are for amplitude of sine wave input > 5% of range. 0.1% errors will be added when the
range of input sine wave is 1% to 5%.
[5]
30 seconds OFF after 30 seconds ON is recommend for the continuous current that higher than DC 7 A or
AC RMS 7 A.
[6]
Specifications are for amplitude of sine wave input > 0.5% of range. For inputs from 1% to 5% of range,
add 0.1% of range extra error.
[7]
Specifications are for 4–wire ohms function or 2–wire ohms using the relative operation of math. Without
relative operation, add ±0.20 Ω additional error in 2-wire ohms function.
[8]
Except for special marks, when frequency ≤ 10M Hz, the specification is applicable to >1V AC input voltage,
when frequency > 10 MHz, the specification is applicable to >3V AC input voltage.
[9]
Specifications are for using the relative operation of math. Using of non-film capacitor may generate
additional errors. Specifications are for from 5% to 110% on ranges.
General Specifications
Display Screen
3.7-inch TFT LCD with resolution 480*320
Operating
Environment
Full accuracy from 0℃ to 50℃, 80% RH and 40℃, non condensing
Storage Temperature: -20℃ to 70℃
Electromagnetic
compatibility
Conforming to EMC Directive 2014/30/EU and EN 55032:2012/AC: 2013
Safety
Conforming to EN 61010-1:2010 and low voltage instructions LVD Directive
2014/35/EU
Remote Interface
RS232
Programmer
Language
Standard SCPI, compatible with commands of main stream multimeters
Warm Up Time
30 minutes
Dimension (W×H×D)
235 x 110 x 295 (mm)
Weight
Approx. 2.3 kg