Digital Storage Oscilloscope 40
2.8.2
Holdoff
Holdoff:
To use Trigger Holdoff, push the HORIZONTAL Menu button and set the
Holdoff Time option by pressing F4. The Trigger Holdoff function can be used to
generate a stable display of complex waveforms (such as pulse trains). Holdoff is the
time between when the oscilloscope detects one trigger and when it is ready to detect
another. During the holdoff time, the oscilloscope will not trigger. For a pulse train, the
holdoff time can be adjusted to let the oscilloscope trigger only on the first pulse in the
train.
2.9 SAVE/RECALL
Oscilloscope setups, waveforms, Ref, and CSV files can be saved to internal
oscilloscope memory or to a USB storage device. The saved setups, waveforms and
Ref can be recalled from an USB Host interface on the front panel to connect an USB
device for external storage.
The scope supports setups, waveforms, Ref and CSV files storage. The default save
type is setups.
1. Setups
It‘s the default storage type of the scope. It saves the settings of the oscilloscope in
i
nternal or external memory in “.SET” format. At most 9 setting files (from No.1
~
No.9)
can be stored in internal memory. The stored settings can be recalled.
2. Wave(Binary)
Trigger Level
Indicates Trigger
Points
Holdoff
Holdoff
Acquisition Interval
Acquisition Interval