3.2 FUNCTIONS OF CONTROL SWITCHES 

NO.  SWITCHES 

FUNCTIONS 

1  POWER SWITCH  Power on/off 

2  POWER LIGHT  Lights when power on 

3  INTENSITY  

Controls brightness of display 

4  FOCUS 

After obtaining appropriate brightness with INTENSITY, 

adjust FOCUS for clearest line 

5  CALIBRATION  Provide symmetric square wave for 0.5V range, 

frequency=1KHz.  Used for adjusting 10:1 Capacitor and 

adjusting the vertical & horizontal sensitivity.  

6  TIME/DIV 

Selects the sweep rate  

7  Horizontal 

POSITION 

Horizontal positioning control of trace on the screen 

8  HORIZONTAL 

VAR 

Provides continuously variable sweep rate, turn clock wise  

to the end is the calibrating position 

9  LEVEL 

Control signal trigger to sweep at certain level 

10  +/-  

TV 

+: Triggering occurs when trigger signal crosses trigger    

  level in a positive-going direction.  

-: Triggering occurs when trigger signal crosses trigger  

  level in a negative-going direction  

TV: used to show TV signals 

11  AUTO/NORM 

EXT/X: X-Y 

AUTO : a single trace shown on screen even no signal.  

       Automatically reverts to triggered sweep  

       operation when adequate triggered signal is  

       present.  Needs to adjust the Level. 

NORM: No trace on screen if no signal. Trace is only  

       generated when adequate trigger signal is present 

EXT/X: X-Y switch 

12  INT/EXT/LINE   Switch to select the Trigger Source INT/EXT/LINE  

13  Ext. Trig Input 

Terminal 

When switching [11] to EXT/X, it’s X-Y input terminal; 

When switching [12] to EXT, it’s Ext. Trig. Input terminal. 

14  ATT 

Continuously adjusting X-Y mode sensitivity, turn 

clockwise to obtain the highest sensitivity 

15  VOLTS/DIV 

Adjusting sensitivity of vertical system 

16  Vertical POSITION Control vertical position of trace on the screen 

17  VERTICAL VAR  Continuously adjusting the sensitivity of vertical system, 

turn clockwise to the end is the calibration position 

18  coupling options 

(AC

DC) 

Selects input coupling options 

19  INPUT 

Vertical input terminal 

20  POWER INPUT 

CONNECTOR 

AND FUSE 

power input connector (refer to the rear panel for voltage) 

110V

±

10% , 220 V 

±

10%;  

 

Содержание AO-610

Страница 1: ...Voltcraft AO 610 10MHZ OSCILLOSCOPE USER S MANUAL ...

Страница 2: ...YSICAL FEATURES 2 9 WORKING ENVIRONMENT 2 10 PRESSURE PROOF TEST 3 CONTROLAND INDICATORS 3 1 CONTROL PANEL POSITION 3 2 FUNCTIONS OF CONTROL SWITCHES 3 3 OPERATING INSTRUCTIONS 4 MEASUREMENT 4 1 EXAMINATION AND ADJUSTMENT BEFORE TAKING MEASUREMENTS 4 2 MEASUREMENT OF VOLTAGE 4 3 TIME MEASUREMENTS 4 4 PHASE DIFFERENCE MEASUREMENTS 4 5 X Y MODE APPLICATIONS 5 ACCESSORIES ...

Страница 3: ...afety and operating instructions in this manual If the instrument is not used as described in this manual the safety features might be impaired WARNING Non compliance with the warnings and or the instructions for use may damage the instrument and or its components or injure the operator Take extreme care under the following conditions when using the instrument For your own safety and that of the i...

Страница 4: ... of any foreign matter from the ventilation holes could result in fire electrical shock or power failure Use this instrument with the rated AC power source Use of this instrument with a voltage other than specified could result in electrical shock fire or power failure The usable power voltage range is marked on the rear panel Do not remove either the cover or panel Take sufficient care when measu...

Страница 5: ... 10DIV 1DIV 6mm Accelerating Voltage 1200V Display Color Green 2 7 POWER SOURCE Voltage Range 110V 10 220V 10 Frequency 50Hz 2Hz Power Consumption 25W 2 8 PHYSICAL FEATURES Weight 3kg Dimensions H x W x D 190 130 270mm 2 9 WORKING ENVIRONMENT Working temperature 5 C 40 C Storage Environment 30 C 60 C 10 80 RH Working Altitude 2000m 2 10 PRESSURE PROOF TEST Pressure proof test 1500V 1min ...

Страница 6: ...5 3 CONTROLAND INDICATORS 3 1 Control Panel Position 3 1 1 Front Panel Fig 3 1 3 1 2 Rear Panel Fig 3 2 ...

Страница 7: ...V used to show TV signals 11 AUTO NORM EXT X X Y AUTO a single trace shown on screen even no signal Automatically reverts to triggered sweep operation when adequate triggered signal is present Needs to adjust the Level NORM No trace on screen if no signal Trace is only generated when adequate trigger signal is present EXT X X Y switch 12 INT EXT LINE Switch to select the Trigger Source INT EXT LIN...

Страница 8: ...M 11 Auto TIME DIV 6 0 5mS SLOPE 10 INT EXT LINE 12 Int AC DC 18 DC 2 OPERATION a Power on 1 b Power lights at 2 c Allow a warm up period of 5 minutes then adjust the intensity 3 d Adjust focus 4 for clearest line If unstable adjust level 9 3 Horizontal Level Adjustment a Slightly turn screws 21 23 counter clockwise but do not remove b Insert a straight end screw driver to 22 and whilst observing ...

Страница 9: ...en 11 set at EXT X The oscilloscope is used for X Y operation at this moment Input 19 is Y axis with the same sensitivity Input 13 as X axis 14 can be adjusted continuously within 0 2V DIV 0 5V DIV 3 3 4 TRIGGER SOURCE In Fig 3 1 12 provides 3 sources for selection INT trigger EXT trigger input from 13 LINE input from power source 3 3 5 HORIZONTAL SYSTEM OPERATION 1 Sweep setup turn the sweep swit...

Страница 10: ...REMENT In order to ensure accuracy and prevent errors the following observations should be made before taking a measurement 4 1 1 TRACE ROTATION The horizontal trace on the screen should be parallel with the horizontal line Due to the earth s magnetic field and other factors which causes horizontal trace leaning you must examine the following before taking measurements 1 Adjust the front panel con...

Страница 11: ...g 4 1 LF COMP Fig 4 2 4 2 MEASUREMENT 4 2 1P P Voltage Measurements Step 1 Input signal to INPUT 19 terminal 2 Setup VOLTS DIV and observe waveform set waveform display on the screen within 5 divisions and turn VAR clockwise to the calibration position 3 Adjust level to make waveform steady 4 Adjust sweep controls to show at least one cycle of the waveform is on the screen ...

Страница 12: ...sweep baseline on the screen 2 Setup input coupling options as 3 Setup POSITION let sweep baseline to coincide with horizontal center define it as the zero reference level 4 Input signal into terminal 5 Set input coupling to DC adjust VOLTS DIV so that the waveform is shown centrally on the screen turn VAR to the calibration position 6 Read the divisions between the zero reference level to the wav...

Страница 13: ...ols to obtain a normal display of 1 2 signal cycles 4 Using the vertical and horizontal positions set two points to be measured in the waveform on the same horizontal level 5 Measure the distance between the two points the measurement is calculated by the following equation Distance between 2 points DIV rate of sweep t Div TIME S Horizontal factor Shown in figure 4 6 distance between A B is 8 Div ...

Страница 14: ... the tops of the waveform coincide with 100 point while the bottoms of the waveform coincide with 0 point 4 Adjust Sweep switch to obtain the positive going direction or negative going direction of the waveform on the screen 5 Use the horizontal POSITION control to adjust the 10 points to coincide with a vertical reference line 6 Measure the horizontal distance in divisions between the 10 and 90 p...

Страница 15: ... X Y mode applications There are some cases which X axis requires control from external signals e g external connection of sweep signals signals of Lissajous patterns or other equipment s display setup X Y mode operation is turn 11 to EXT X input signals through 13 sensitivity to be adjusted directly with 14 then input Y signal through 19 5 Scope of Delivery 1 Oscilloscope 2 Cord 1 pc 3 User s man...

Страница 16: ...15 ...

Отзывы: