83
Specified measuring range: 5–100% of the measuring range
600 V overload protection; Impedance: 400 kΩ (max. 250 V, 3 secs)
The multimeter may display ≤5 counts if a measuring input is short-circuited.
After using the LoZ feature, leave the multimeter for 1 minute before using it
again.
3
4
Alternating voltage (V/AC)
Range
Resolution
Accuracy
2.000 V
0.001 V
±(0.9% + 9)
20.00 V
0.01 V
±(1.3% + 7)
200.0 V
0.1 V
600 V
1 V
Specified measuring range: 5–100% of the measuring range
Frequency range: 45–400 Hz; 600 V overload protection; Impedance: 10 MΩ
The multimeter may display 5 counts if a measuring input is short-circuited.
TrueRMS peak (Crest Factor (CF)) ≤3 CF to 600 V
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
5
Alternating voltage (V/AC) LoZ
Range
Resolution
Accuracy
2.000 V
0.001 V
±(1.9% + 5)
20.00 V
0.01 V
200.0 V
0.1 V
600 V