Measuring tolerances
Accuracy in ± (% of r display error in counts (= number of smallest points)).
These accuracy readings are valid for one year at a temperature of +23 °C (± 5
°C) and a relative humidity of less than 75 % (non-condensing). If the multimeter is
used outside of this temperature range, use the following coefficient to calculate the
ac0.1 x (specified accuracy)/1 °C.
The accuracy of measurements may be affected when the multimeter is used in a
high-frequency electromagnetic field.
1
Direct voltage (V/DC)
Range
Resolution
Accuracy
200.0 mV
0.1 mV
±(0.9% + 8)
2.000 V
0.001 V
±(0.9% + 4)
20.00 V
0.01 V
200.0 V
0.1 V
600 V
1 V
±(1.3%+7)
Specified measuring range: 5–100% of the measuring range
600 V overload protection; Impedance: 10 MΩ (mV: ≤100 MΩ)
The multimeter may display ≤5 counts if a measuring input is short-circuited.
2
Direct voltage (V/DC) LoZ
Range
Resolution
Accuracy
2.000 V
0.001 V
±(1.7% + 7)
20.00 V
0.01 V
200.0 V
0.1 V
600 V
1 V