10. Source (output mode)
If the calibrator is in “Measurement Mode”, press
to switch to “Output Mode” and the LCD will display
“SOURCE”.
a) Millivolt
Press
to select millivolt output. Press
to choose output digit, then press
to set the value.
The figures below show millivolt output and how to connect the test leads.
Millivolt output
Connection
b) Voltage
Press to select voltage output. Press
to choose output digit, then press
to set the value.
The figures below show voltage output and how to connect the test leads.
Voltage output
Connection
c) Thermocouple simulation
The calibrator can measure 10 models of thermocouples, including R, S, K, E, J, T, N, B, L, U.
1. Press
. The thermocouple reading will be displayed.
2. Press
one more time to select appropriate model of thermocouple.
3. Press
to set the manual compensation value of cold junction temperature, then press arrow keys to adjust
the manual compensation value.
4. Press
to finish.
5. Short press
to switch to automatic compensation.
Refer to the figure below for connecting the test leads.