9. Advanced functions
a) Setting user-defined values for 0% and 100%
When setting user-defined values as 100% or 0%, you can use the “SET 100%” and “SET0%”
buttons to save your own values. These values are retained in the device when it is turned off.
Set the desired voltage value and then press and hold the “SET100%” or “SET0%” button for
>1.5 s.
Each setting to 100% or 0% has different threshold values depending on the range. If
a value cannot be set, select a higher value (for 100%) or a lower value (for 0%).
The user-defined values can be reset to their factory defaults. To do this, turn off the device.
Press and hold the “RTD/ESC” button and turn on the device. Press and hold both buttons until
you hear a beep. The factory settings have been restored.
The table below shows the factory defaults:
Output function
0% value
100% value
mV (100 mV)
0.000 mV
100.000 mV
mV (1000 mV)
0.0 mV
1000.0 mV
Resistance 500 Ω
0.0 Ω
500.0 Ω
Resistance 5000 Ω
0 Ω
5000 Ω
Type J
0.0 °C
1000.0 °C
Type K
0.0 °C
1000.0 °C
Type T
0.0 °C
400.0 °C
Type E
0.0 °C
800.0 °C
Type R
0 °C
1500 °C
Type S
0 °C
1500 °C
Type B
600 °C
1800 °C
Type N
0.0 °C
1000.0 °C
48