3.Appendix

 

 

15 

 

Appendix C: Line Fuse Replacement   

The line fuse is in the plastic fuse box below the power line input on the rear panel. 

 

 

Warning: 

Disconnect the line cord at the rear panel and remove all test leads 

connected to the instrument before replacing the line fuse. Failure 

to do so could expose the operator to hazardous voltages that could 

result in personal injury or death. 

Use only the correct fuse type. Failure to do so could result in personal injury or 

instrument damage. 

Voltage 

Fuse 

90 - 240 V AC 

250 V, F1AL 

To perform the line fuse replacement, follow these steps: 
1.  Turn off the multimeter, remove all measurement leads and other cables from the 

instrument, including the power cord. 

 

2.  Use a flat-blade screwdriver to remove the fuse box. 

 

 

 

 

 

 

 

 

3.  Replace the fuse with a new one, which should match with the rating; install it into 

the fuse box, and push the fuse box back on to the rear panel. 

 

 

 

 

 

 

 
 

2019.09  V1.0.0

 

 

 

   

 

Содержание 2203066

Страница 1: ...VC 7055BT Digital Multimeter Quick Guide...

Страница 2: ...erminals HI Sense and LO Sense Measurement Limits 4 Measurement Category 4 2 Quick Start 5 General Inspection 5 Dimensions 5 Foot Stool Adjustment 6 Front Panel Overview 6 Rear Panel Overview 8 User I...

Страница 3: ...e instrument or other property might occur Safety Symbols Symbols on the Product The following symbol may appear on the product Direct current DC Warning risk of electric shock Alternating current AC...

Страница 4: ...s product Refer to the user s manual for the Measurement Limits before connecting to the instrument Do not exceed any of the Measurement Limits defined in the following section Do not operate without...

Страница 5: ...equency period capacitance diode and temperature test measurements Two Measurement Limits are defined for these terminals HI Input to LO Input Measurement Limit The Measurement Limit from HI Input to...

Страница 6: ...ory II Inputs may be connected to AC mains power up to 600 VAC under Category II overvoltage conditions Measurement category definition Measurement CAT I applies to measurements performed on circuits...

Страница 7: ...l You can check whether there is any loss of accessories with reference to this description If it is found that there is any accessory lost or damaged please get in touch with our distributor responsi...

Страница 8: ...w 2 3 4 5 12 1 6 7 8 11 9 10 Figure 2 1 Front panel overview Item Name Description 1 LCD Display the user interface 2 Menu selection Keys Activate the corresponding menu 3 Operation Keys Save Collect...

Страница 9: ...asurements 6 Range Direction Keys When the Range softkey is shown on the right menu you can press the key to switch between auto and manual range Press to enable manual range and increase or decrease...

Страница 10: ...ng will be displayed in the Vice Display 12 Dual Exit Press the key to exit dual display mode Rear Panel Overview 1 3 2 5 4 Figure 2 2 Rear panel overview Item Name Description 1 RS232 Connect the PC...

Страница 11: ...ay Power On 1 Connect the instrument to the 90 240 V AC supply using the supplied power cord Warning To avoid electric shock the instrument must be grounded properly 2 Press down the power button at t...

Страница 12: ...trument damage do not discretionarily switch the measurement function when measuring DC Voltage Measurement DC Voltage AC Voltage Measurement AC Voltage DC Current Measurement DC Current AC Current Me...

Страница 13: ...anel key to save current reading to the data table of the internal memory Auto record After setting number of points sample interval press the Start softkey to start recording After recording you can...

Страница 14: ...ord 1 Configure the parameters Press the front panel key press the Auto softkey Press the Point softkey to specify the total number of readings to record The range is 1 to 1000 Press the Interval soft...

Страница 15: ...l exists please contact us for our service 2 The reading does not change when a current signal is input 1 Check whether the test lead is correctly inserted into the current input terminals I terminal...

Страница 16: ...prevent electrical shock disconnect the instrument from AC mains power and disconnect all test leads before cleaning 2 Clean the outside of the instrument using a wet soft cloth not dripping water Do...

Страница 17: ...in personal injury or death Use only the correct fuse type Failure to do so could result in personal injury or instrument damage Voltage Fuse 90 240 V AC 250 V F1AL To perform the line fuse replacemen...

Отзывы: