Title: Instruction manual PROFINET interface
Date: Sept. 2020
Ver: 04AD_Red-y_V2.1
By: WES/HAE
Page 13
/
30
Process Data Objects (PDO’s)
PDO’s are variables continuously transferred between the Device (slave) and the PLC (master)
The PROFINET interface supports only one slot for input and one slot for output.
Each sub-slot can have one register (from the list below) assigned to it. The input slot supports a maximum
of 20 sub-slots. The output slot supports 12 sub-slots. When adding a sub-slot, a list is available (through the
GSDML) with all the available registers. The order in which the registers can be added is not fixed.
Appendix A shows an example in TwinCAT.
This document describes how the data is encoded in the PDO’s. The name of the register corresponds to the
name used in the GSDML. The tables also show the Modbus registers which are linked to the PDO data.
More information about the registers can be found on the Vögtlin website under the download section, look
for digital communication.
Read
with
Input PDO’s (slave to master)
Register
Modbus Addr.
(zero base)
Data
Type
Description
Flow
0x0000
F32
Actual flow
Temp
0x0002
F32
Gas Temperature
Totalizer
0x0004
F32
Accumulated gas total
Set point Flow
0x0006
F32
Set point flow when in controller mode
Valve Power
0x000A
F32
Read valve PWM in %
Alarms
0x000C
UINT16 Indicates the alarm messages in a bit map
HW Status
0x000D
UINT16 Hardware error status register
Device setup
0x000E
UINT16 Setup control mode
Ramp Time
0x000F
UINT16 Changing time that it takes between set points
Flow Unit
0x0016
STR8
Flow unit
Gas Name
0x001A
STR8
Name of current gas
Serial Number
0x001E
UINT32 Serial number of the device
Device Type1
0x0023
STR8
Name of the instrument type / instrument code
PID Select
0x0035
UINT16 Select a PID preset for flow controller
Flow Limit
0x094F
F32
Maximum flow allowed
Device Type2
0x1004
STR8
Name of the instrument type / instrument code
Totalizer Unit
0x4048
STR8
Totalizer Unit
Enable SP storage
0x4050
UINT16 Enable the storing of the set point in EEPROM
LUT Select
0x4139
UINT16 Select a LUT from the calibrated list
Pressure
0x5F00
F32
Actual pressure (only with pressure Transmitter)
Pressure Min
0x5F02
F32
Minimum pressure set point
Pressure Max
0x5F04
F32
Maximum pressure set point
Set point pressure
0x5F06
F32
Set point pressure when in controller mode
Pressure Unit
0x5F08
STR8
Pressure Unit
Pressure PID select
0x5F10
UINT16 Select a PID preset for pressure controller