ViZion DR + Straight Arm Installation Guide v2.0
Opal UAI 2.3.x.x
2
Customer Support
1.800.366.5343 –
VZDR-UG-ViZion DR + Straight Arm Installation Guide-D-16082013.2.0
ViZion DR + Straight Arm Installation
This guide will provide you with detailed instructions on how to assemble and mount the ViZion DR +
within the Viztek Straight Arm.
1.
Remove the following:
Figure 1
a.
Grid holder
b.
Bucky cover
c.
Bucky front plate
2.
Remove the 12 screws fastened to the bottom of the bucky cover.
3.
Remove the 4 screws that hold the AEC/grid guide to the bucky.
4.
Attach ViZion DR+ to the mounting plate, using 2 screws on each side.
Figure 2
NOTE: Install a grounding strap to the ViZion DR+
detector in the screw hole labeled with the
universal grounding symbol.
5.
Place the ViZion DR + dector and the mounting plate
onto the bucky
6.
Align the plate with the same screw holes used for the
AEC/grid guide which were removed in step 3.
7.
Place the AEC/grid guide on top of the panel and fasten 4 machine screws from the bottom.
8.
Plug the correct cables to the ViZion DR+ detector plate and the AEC/grid.
9.
Mount bottom plate attaching the grounding strap from ViZion DR + detector to the universal
ground located on the bottom cover.
Grid Holder
Bucky Front Plate
Bucky Cover
Figure 1
Figure 2